6.4 COMMAND OVERVIEW
Data Byte section of the TCP/IP Datagram
Command Name
<CMD>
<ARG>
RANGE
Program RS-232
unit baud rate
07
1 ASCII
1 - 5
Program User
Configs
09
9 ASCII
See
Description
Program kV
10
1-4 ASCII
0-4095
Program mA
11
1-4 ASCII
0-4095
Request kV
Setpoint
14
None
-
Request mA
Setpoint
15
None
-
Request Analog
Monitor
Readbacks
19
None
-
Request HV On
Hours Counter
21
None
-
Request Status
22
None
-
Request Software
Version
23
None
-
Request Hardware
Version
24
None
-
Request Web
Server Version
25
None
-
Request Model
Number
26
None
-
Reset HV On
Hours Counter
30
None
-
Reset Faults
31
None
-
Read Interlock
Status
55
None
-
Request kV
monitor
60
None
-
Request mA
monitor
61
None
-
Request –15V
LVPS
65
None
-
Wathdog Tickle
88
None
-
Wathdog enable
89
1 ASCII
0 or 1
Turn HV on/off
98
1 ASCII
0 or 1
118080-001 REV A
Page 69 of 95
Summary of Contents for SLM SERIES
Page 15: ...SLM MANUAL 6 118073 001 Rev C Figure 2 2 Unit Dimensions 1200W ...
Page 19: ...SLM MANUAL 10 118073 001 Rev C Figure 3 3 Local Programming via External Voltage Source ...
Page 20: ...SLM MANUAL 11 118073 001 Rev C Figure 3 4 Remote Monitoring ...
Page 40: ...Figure 9 Web Page 1 Contact Information 118080 001 REV A Page 13 of 95 ...
Page 98: ...Request Faults 68 20 ASCII 118080 001 REV A Page 71 of 95 ...