
The format of Data Bytes 1 through N are as follows:
<STX><CMD><,>ARG><,><ETX>
Where:
<STX>
= 1 ASCII 0x02 Start of Text character
<CMD>
= 2 ASCII characters representing the command ID
<,>
= 1 ASCII 0x2C character
<ARG>
= Command Argument
<,>
= 1 ASCII 0x2C character
<ETX>
= 1 ASCII 0x03 End of Text character
5.2 COMMAND
ARGUMENTS
The format of the numbers is a variable length string. To represent the number
42, the string ‘42’, ‘042’, or ‘0042’ can be used. This being the case, commands
and responses that carry data are variable in length.
5.3 COMMAND
OVERVIEW
Data Byte section of the TCP/IP Datagram
Command Name
<CMD>
<ARG>
RANGE
Program RS-232
unit baud rate
07
1 ASCII
1 - 5
Program User
Configs
09
9 ASCII
See
Description
Program kV
10
1-4 ASCII
0-4095
Program mA
11
1-4 ASCII
0-4095
Request kV
Setpoint
14
None
-
Request mA
Setpoint
15
None
-
Request Analog
Monitor Readbacks
19
None
-
Request HV On
Hours Counter
21
None
-
Request Status
22
None
-
Request Software
Version
23
None
-
Request Hardware
Version
24
None
-
Request Web
Server Version
25
None
-
Request Model
Number
26
None
-
118080-001 REV A
Page 38 of 95
Summary of Contents for SLM SERIES
Page 15: ...SLM MANUAL 6 118073 001 Rev C Figure 2 2 Unit Dimensions 1200W ...
Page 19: ...SLM MANUAL 10 118073 001 Rev C Figure 3 3 Local Programming via External Voltage Source ...
Page 20: ...SLM MANUAL 11 118073 001 Rev C Figure 3 4 Remote Monitoring ...
Page 40: ...Figure 9 Web Page 1 Contact Information 118080 001 REV A Page 13 of 95 ...
Page 98: ...Request Faults 68 20 ASCII 118080 001 REV A Page 71 of 95 ...