DO NOT CHANGE THESE SETTINGS: they are factory
configured.
7) Oven set point
Defines the oven temperature in degrees Celsius and is used by the oven
temperature control loop.
The real temperature of the oven is displayed in the third column of the Oven Set
Point grid. To change the set point, change the value in the second column for
the corresponding oven. Be aware that the desired set point can take several
minutes to be reached and stabilised.
Minimum
0
Maximum
300
Resolution
1
Regen
200
Table 19 : Oven set point
DO NOT CHANGE THESE SETTINGS: they are facto
configured.
8) Remote Starting Enable
When this check box is checked, a cycle can be started by the digital input #1.
The edit box specifies the time that elapsed before the cycle starts after the digital
input #1 is activated. When the 24 Volts is applied to the digital input #1, the
remote starting countdown starts (it can be viewed in the
RUN-REAL TIME
CHROMATOGRAM MENU
). Remove the voltage when the countdown
reaches 0 to start a cycle, or else the countdown will restart.
9) Screen Saver Enable
When this check box is checked, the screen saver is enabled. The edit box
specified the time that elapsed before the screen power off. By moving the mouse
or pressing a key, the screen will power on.