SERVOPRO CHROMA Gas Analyser
60
04400001A / Revision 7
4.6 Configuration Menu
4.6.1
System
Press
CTRL-S
or click
Configuration
on the menu bar and then
System
to
access the
CONFIGURATION-SYSTEM
menu (
). The
System
configuration menu allows adjustment of several key parameters.
Figure 22 : Configuration - System menu
1) Analogue output
Two modes are available. The
Hold
mode maintains the 4-20 mA outputs even
if a new cycle is started. The
Track
mode ensures the analogue outputs always
track input gas values.
2) mA Failure mode
In the case of a system status alarm, this control sets all the 4-20 mA outputs
below 4mA if
LOW
is selected and higher than 20 mA if
HIGH
is selected.
Selecting
OFF
will disable this function and the 4-20 mA outputs reflect signal
values as in normal operation.