Modbus Communication
OXY5500 Operator’s Manual
4
–
11
Table 4-1
Holding registers (Continued)
Register
Name
Address
Size
Variable
Type
Description
Write
Access
Subnet Mask
5685
8
Int32
The Subnet Mask. Each
pair of registers holds one
octet of the Address. Refer
to
details. This register will
only be used if register
5675 is set to “0” (DCHP
off).
Yes
Ethernet Port
for Modbus
5693
2
Int32
The Ethernet Port used in
the Modbus Protocol.
(Default: 502)
Yes
Ethernet
Modbus ID
5695
2
Int32
The Ethernet Modbus ID
(range: 0-32).
Yes
Alarm Relay
High Level
5697
2
Float
The High Level which trig-
gers the Level Alarm Relay.
The unit used is the same
as defined in Table 4–14.
Yes
Alarm Relay
Level Low
5699
2
Float
The Low Level which trig-
gers the Level Alarm Relay.
The unit used is the same
as defined in Table 4–14.
Yes
Pressure
Mode
5705
1
Boolean
Sets the measurement
mode to either acquisition
via 4-20mA or fixed value
where:
0x00 = fixed value
0x01 = 4-20mA
Yes
Measure-
ment Mode
5707
2
Int32
This is a bit-coded register
to configure the measure-
ment mode and to trigger
the measurement start.
•
Bit 0
: Reserved.
•
Bit 1
: Read only. Set
when a measurement
is already active.
•
Bit 2
: Will perform a
single scan.
Yes
Set Concen-
tration Alarm
Low Level
5709
2
Int32
Enables/disables the Low
Level Alarm of the Concen-
tration Alarm Relay:
0x00: Disable (Low Level is
ignored
0x01: Enable
Yes
Summary of Contents for OXY5500
Page 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Page 2: ......
Page 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...