Modbus Communication
OXY5500 Operator’s Manual
4
–
21
Table 4–21
Measurement reading for calibration process example
Register
4899/4900
Register
4901/4902
Register
4903/4904
Register
4905/4906
Register
4907/4908
Float:
Oxygen
Amplitude in
V
Float:
Oxygen
Phase shift in
degrees
Float:
Temperature
in
C
Float:
Calculated Oxy-
gen Value in unit
Int32: Error
Register.
Refer to
Table 4–10.
50592.62
(Sensor and
environment
dependent
value)
66.32
(Sensor and
environment
dependent
value)
21.98
This value can
be ignored while
calibration
process takes
place.
0 (Error
code.
Should be 0
if a sensor is
connected)