Troubleshooting & Maintenance
OXY5500 Operator’s Manual
B
–
11
Removing and Replacing the Oxygen Probe
Use the following instructions to remove and replace an oxygen probe on the
OXY5500.
Tools/Parts
•
OXY5500 replacement oxygen probe
•
Adjustable Crescent wrench
•
Phillips screwdriver
•
5/32” Hex driver
•
7/16” Open-end wrench
•
1/2” Open-end wrench
Removing the Oxygen Probe
Use the following steps to remove a faulty oxygen probe.
1.
Purge the analyzer by allowing 99.9999% pure Nitrogen flow
through the system for 30 minutes.
2.
Shut off gas flow to the analyzer.
3.
Turn off power to the analyzer.
4.
Loosen the enclosure screws and remove the clamps to open the
enclosure door.
5.
Using an adjustable wrench, loosen the cable gland cap
on the panel
by turning ‘up’ towards the analyzer. Do not remove cable gland cap.
Refer to Figure B–10.
on page A-6 for a complete list of
replaceable probe parts and part numbers.
Summary of Contents for OXY5500
Page 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Page 2: ......
Page 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...