OXY5500 Operator’s Manual
A
–
1
Appendix A: Specifications
Table A–1
OXY5500 analyzer specifications
Application Data
Target Components
O
2
Principle of Measurement
Fluorescent Quenching
Typical Measurement Ranges
OP-9
1
1. Analyzer ships with one sensor. Additional sensors are optional.
OP-6
1
OP-3
1
0-200 ppmv (default)
0-10 to 10-1,000 ppmv
2
User setting
2. Accuracy specification applies to concentrations up to 300 ppmv.
0-5%
0-1 to 0-5%
User setting
0-50%
0-10 to 0-100%
User setting
Lower Limit of Detection
0.5 ppmv
20 ppmv
300 ppmv
Accuracy at 20-25
C
5% of Reading
3% of Reading
2% of Reading
Repeatability
1% of Reading
Measurement Update Time
Programmable Sampling Rate (default 30 seconds)
Temperature Range
-20
C to 60
C (-4
to 140
F)
Sample Inlet Pressure
140-275 KPaG (20-40 PSIG) to sample panel regulator
Sample Pressure Range
800-1400 mbara
Maximum Probe Pressure
275 KPaG (40 PSIG)
Sample Flow Rate
Typical 1.0 SLPM (2.1 SCFH)
Recommended Calibration
Two-point calibration in oxygen-free environment (Nitrogen)
and a second span point (cylinder gas). Validate with O
2
in N
2
reference (cylinder gas).
Electrical & Communications
Input Power (Voltage and
Max. Power)
100-240 VAC, 50/60 Hz or 9-36 VDC
5 Watts @ 24 VDC, 14 Watts @ 120 VAC, 22 Watts at 240 VAC
Communication
Analog
: Qty (2) 4-20mA sourcing power outputs and (1) 4-20
mA input (sample pressure)
Fieldbus
: RS-232C, RS-485, Ethernet 10/100 with Modbus
Output Relays
: Qty (2) 250 mA max load (Concentration and
Fault Alarms)
USB 2.0 works with Service Software only
4 GB Internal Memory with Internal Data Logging
LCD Display
Concentration, Temperature, Sample Rate, Data Logging, Diag-
nostics, plus Full Menu for Setup, Calibration, etc.
Service Software
Windows software. Connect via USB port. Download data logs,
trend and monitor, calibrate and troubleshoot.
Summary of Contents for OXY5500
Page 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Page 2: ......
Page 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...