Oxygen Analyzer
3
–
12
4900002239 rev. G 7-11-17
The display switches to the main measurement screen automatically. Refer to
Figure 3–1.
To achieve a highest accuracy, the OXY5500 should be warmed up for
approximately five minutes before taking a measurement.
After the warm-up, complete a field calibration to achieve accurate
measurements. Two standards are required; an Ultra High Purity Nitrogen
bottle and a 80-100 ppm oxygen in Nitrogen balance bottle. Refer to
Sensor Menu Options
The option to change parameters, sensor type or to calibrate the analyzer are
accessed via the
Sensor
button on the MAIN MENU.
Change Parameters
Clicking the
Change Parameters
button in the SENSOR menu causes a
message window to display with the question to abort the currently running
measurement. Refer to Figure 3–16.
Warm-up time may be extended up to 15 minutes if the optode
has been exposed to high concentrations of oxygen.
Figure 3–16
Message window - Stop
measurements during configurations
Summary of Contents for OXY5500
Page 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Page 2: ......
Page 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...