Operation
OXY5500 Operator’s Manual
3
–
43
•
The
Baudrate
for the RS-232 channel can be set to 9600, 19200,
38400, 57600 or 115200.
•
The
ID
, which will be used within the Modbus communication, can be
set to any value between 1 and 32.
•
Parity
can be set to Even, Odd or None.
All settings will be applied by pressing
.
RS-485 settings
Use this screen to set the baudrate of the RS-485 channel. Refer to Figure
•
The
Baudrate
for the RS-485 channel can be set to 9600, 19200,
38400, 57600 or 115200.
•
The
ID
, which will be used within the Modbus communication, can be
set to any value between 1 and 32.
•
Parity
can be set to Even, Odd or None.
Setting the Parity to “None” will also set the number of stop bits
to two. Odd and Even settings will use one stop bit.
Setting the
Parity
to None will also set the number of stop bits to
two. Odd and Even settings will use one stop bit.
Figure 3–47
Digitals - RS-485 settings
Summary of Contents for OXY5500
Page 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Page 2: ......
Page 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...