Operation
OXY5500 Operator’s Manual
3
–
25
At the top of the screen, the currently measured Oxygen, Temperature, and
Pressure values are displayed. Below this, the Old RATA Mult. value is shown.
Begin by setting the reference values.
1.
Enter the reference oxygen value (oxygen concentration of certified
test gas lead into the vessel with the oxygen sensor, or the oxygen
value of a reference device) in the
O2 Reference
field (1) at the
bottom of the screen.
2.
Press the
Start
button next to the
New RATA Mult.
field (2), as
shown in the
Status
field, to display current sensor phase values.
Wait while the sensor values stabilize until the
Status
field displays
“Ready to Set Value!” (3).
3.
Press the
Set
button (4) next to the
New RATA Mult.
field and the
new value will display.
The New RATA Mult. can also be set manually. Refer to
4.
Press the
Save
button at the upper right of the screen.
If RATA has not been changed, the display will read 1.000.
Figure 3–30
Relative Accuracy Test Audit (RATA)
Screen
2
1
3
4
Summary of Contents for OXY5500
Page 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Page 2: ......
Page 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...