Operation
OXY5500 Operator’s Manual
3
–
13
Select
Yes
to stop the measurement in order to display the SENSOR TYPE AND
SENSOR CONSTANTS window. Refer to Figure 3–17.
Use the arrow buttons to navigate between input fields.
To enter editing mode:
1.
Press .
a.
Change the setting or value (one digit at a time) by pressing the
buttons.
b.
Make the desired change to an input field.
2.
Press
again to save the changes.
To exit editing mode:
1.
Press
to cancel and exit.
Changing the Sensor Type
Change the Sensor Type (
OP-3
,
OP-6
or
OP-9
) according to the sensor that
is connected to the analyzer. The displayed sensor constants (
dKSV1
,
dKSV2
,
dPhi1
,
dPhi2
,
f1
and
m
) will change according to the sensor type selected.
The sensor constant values can also be located on the Calibration
Certificate delivered with the optical oxygen sensor. Refer to the
Figure 3–17
Sensor Type and Sensor Constants
- Sensor Type menu selected for editing
Summary of Contents for OXY5500
Page 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Page 2: ......
Page 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...