Sentry 2.0 Controls
VirTis Freezemobile 25L and 35L
16
Rev 005, 02/12
© SP Scientific 2012
Vacuum Value
To adjust the Vacuum value:
1.
Open the condenser door and adjust the trim pot located on the main Sentry
2.0 module within the electrical box.
2.
Set the value to 760 Torr, or current atmospheric pressure.
3.
Turn the trim pot clockwise to increases the pressure or counter-clockwise to
decrease the pressure.
4.
Once the rough vacuum has been set, connect a reference tube.
5.
Adjust the Vacuum value located on the LCD screen up to increase the
Vacuum reading. Adjust the Vacuum value down to decrease the Vacuum
reading.
Note: SP Scientific recommends the use of a Varian Vacuum Reference Tube for
vacuum calibration.
PRACTICAL OPERATING TIP. ENSURE THAT THE GROUND WIRE IS FIRMLY CONNECTED
BEFORE CALIBRATING ANY VALUES. A LOOSE GROUND WIRE CAN PRODUCE
INCONSISTENT AND UNPREDICTABLE RESULTS.
Turn vacuum trim pot here
Summary of Contents for Freezemobile Series
Page 1: ...FREEZE DRYER OPERATOR S MANUAL VIRTIS FREEZEMOBILE 25L AND 35L FREEZE DRYERS...
Page 2: ......
Page 6: ...iv Rev 005 02 12 SP Scientific 2012...
Page 10: ......
Page 12: ...Introduction VirTis Freezemobile 25L and 35L 2 Rev 005 02 12 SP Scientific 2012...
Page 16: ...Installation and Startup VirTis Freezemobile 25L and 35L 6 Rev 005 02 12 SP Scientific 2012...
Page 32: ...Basic Operation VirTis Freezemobile 25L and 35L 22 Rev 005 02 12 SP Scientific 2012...
Page 40: ...Optional Components VirTis Freezemobile 25L and 35L 30 Rev 005 02 12 SP Scientific 2012...
Page 46: ...General Maintenance VirTis Freezemobile 25L and 35L 36 Rev 005 02 12 SP Scientific 2012...
Page 50: ...Appendix A Troubleshooting VirTis Freezemobile 25L and 35L 40 Rev 005 02 12 SP Scientific 2012...
Page 53: ......