VirTis Freezemobile 25L and 35L
Installation and Startup
Rev 005, 02/12
5
© SP Scientific 2012
5.
Open the condenser door. A very light film of high-vacuum grease has been
applied to the front contact surface of the condenser gasket prior to factory
testing. Only a light sheen should be visible on the surface. Check often to
ensure that this film is present.
Note: Never grease the inside of the gasket where it grips the condenser lip.
6.
Ensure that the voltage, phase, frequency and amperage listed on the
Freezemobile serial tag match the available power supply.
7.
Switch the main power switch to the On position. The Sentry 2.0 display will
illuminate.