12070T
!
Autoranging Multimeter
Hz
Temperature Measurements
1. Set the rotary function switch to the
Temp °F or Temp °C
position.
2. Connect the Temperature Probe to the
Banana Plug Adapter.
Insert the adapter into the negative
COM
and the positive
Temp
input jacks, making
sure to observe the correct polarity.
3. Touch the tip of the Temperature Probe to the
part you wish to measure. Keep the probe
touching the part under test until the reading
stabilizes (about 30 seconds).
4. Read the temperature on the LCD display.
10A
30 sec MAX
CAT III 1000V
CAT IV 600V
FUSED
MAX 400mA
Temp
Auto Power Off
Waterproof
every 15 min
mA
COM
Hz
CAP
®
12070T
!
Autoranging Multimeter
K-T
YP
E
Operation
1. Set the rotary function switch to the
Hz
position.
2. Insert the black test lead into the
negative
COM
input jack and the red
test lead into the positive
Hz
input jack.
3. Touch the test probe tips to the circuit
under test.
4. Read the frequency on the LCD display.
Frequency Measurement
Observe all safety precautions when working on live voltages.
WARNING:
To avoid electric shock, do not let the temperature
probe contact live circuits.
WARNINGS:
10A FUSE
0.5A FUSE
12
Operation
Replacing the Fuses
To avoid electric shock, remove test leads from the meter before removing the fuse cover.
WARNINGS:
When replacing the battery or fuses, be sure to secure the back cover panel and battery
compartment door firmly to maintain the waterproof and dust proof integrity of the meter. Loose or overtightened
screws, or an improperly seated o-ring may compromise the meter's water and dust ingress protection.
WARNINGS:
To avoid electric shock,
do not operate your meter until the battery and
fuse cover are in place and fastened securely.
WARNINGS:
To avoid electric shock, do not operate your meter until the battery door is in place and
fastened securely.
WARNING:
1. Remove the six Phillips screws that secure the
back cover to the meter. (Lift the tilt stand to
expose the two bottom screws.)
2. Remove back cover.
3. Remove the old fuse from its holder
by gently pulling it out.
4. Install the new fuse into the holder.
5. Always use a UL listed fuse of the proper size
and value (0.5A/1000V (6.3 x 32mm) fast blow
for the 400mA range, (10A/1000V (10 x 38mm)
fast blow for the 10A range).
6. Install back cover and tighten screws.
A =
Back Cover Screws
B =
Battery Screws
9V BATTERY
A
A
A
A
B
B
Replacing the Battery
To avoid electric shock, remove test leads from the meter before removing the battery door.
WARNING:
1. When the battery drops below the operating voltage, the symbol will
appear on the LCD display. The battery should be replaced immediately.
2. Remove the two Phillips screws from the
battery door located on the back of the meter.
3. Remove door.
4. Replace battery with a fresh 9V battery.
5. Install battery door and tighten screws.
!
W
ARNING
TO A
VOID
ELECTRICAL
SHOCK,
REMOVE ALL
INPUTS BEFORE
OPEN
ING THE CASE.
TO PR
EVE
NT FIRE
, INST
ALL
FUSE
WITH AMP/VOL
T
RA
TING SHOW
N:
9V BA
TTER
Y
NEDA
1604 IEC 6F22
F10A/
10
00V
F500mA/1000V
When replacing the battery or fuses, be sure to secure the back cover panel and battery
compartment door firmly to maintain the waterproof and dust proof integrity of the meter. Loose or overtightened
screws, or an improperly seated o-ring may compromise the meter's water and dust ingress protection.
WARNING: