Operation

AC and DC Current Measurements cont.

7. Touch the test probe tips in series with the circuit being tested.

    For DC measurements, touch the red test probe to the positive side of

    the circuit and the black test probe to the negative side of the circuit.
8. Apply power to the circuit.
9. Read the current on the LCD display. 

Resistance Measurements

 

               

Never test resistance on a live circuit.

WARNING:

 

1. Set the rotary function switch to the

 Ω

            position. 

2. Press the 

MODE

 button until “

M

Ω

    appears on the LCD display.
3. Insert the black test lead into the negative

   

 COM

 input jack Insert the red test lead into

    the positive 

Ω

 input jack.

4. Touch the test probe tips across the circuit or

    component under test. It is best to disconnect one

    side of the device under test so the rest of the circuit

    will not interfere with the resistance reading.
5. Read the resistance on the LCD display.

10A

30 sec MAX

CAT III 1000V

CAT IV 600V

FUSED

MAX 400mA

Temp

Auto Power Off

Waterproof

every 15 min

mA

COM

Hz

CAP

®

12070T

!

Autoranging Multimeter

Ω

Continuity Test

 

               

Never test continuity on a live circuit.

WARNING:

 

1. Set the rotary function switch to the

    

Ω

           position.

2. Press the 

MODE

 button until the “    ” symbol

    appears on the LCD display. 
3. Insert the black test lead into the

    negative 

COM

 input jack. Insert the red

    test lead into the positive 

Ω

 input jack.

4. Touch the test probe tips to the circuit or

    wire you wish to check. 
5. If the resistance is approximately 30

Ω

 or less,

    an audible tone will sound. If the circuit is above 400

Ω

,

    the LCD display will indicate “

OL

”. The resistance will

    be shown on the LCD display if it is below 400

Ω

10A

30 sec MAX

CAT III 1000V

CAT IV 600V

FUSED

MAX 400mA

Temp

Auto Power Off

Waterproof

every 15 min

mA

COM

Hz

CAP

®

12070T

!

Autoranging Multimeter

Wire

10

Operation

Diode Test

  

 Never test diodes in a live circuit. 

WARNINGS:

 

1. Turn the rotary function switch to

    the 

Ω

           position.

2. Press the 

MODE

 button until the “      ”

    symbol appears on the LCD display.
3. Insert the black test lead into the negative

    

COM

 input jack and the red test lead into

    the positive 

Ω

 input jack.

4. Touch the test probes to the diode under test.

    Forward voltage will indicate 0.4 to 0.7V.

    Reverse voltage will indicate “

OL

”. Shorted

    devices will indicate near 0V and an open

    device will indicate “

OL

” in both polarities.

Red

Probe

Red

Probe

Black

Probe

Black

Probe

Forward test                 Reverse test

10A

30 sec MAX

CAT III 1000V

CAT IV 600V

FUSED

MAX 400mA

Temp

Auto Power Off

Waterproof

every 15 min

mA

COM

Hz

CAP

!

Autoranging Multimeter

10A

30 sec MAX

CAT III 1000V

CAT IV 600V

FUSED

MAX 400mA

Temp

Auto Power Off

Waterproof

every 15 min

mA

COM

Hz

CAP

®

12070T

!

Autoranging Multimeter

WARNING:

 

1. Set the rotary function switch to the 

CAP

 position.

2. Insert the black test lead into the negative

    

COM

 input jack and the red test lead into

    the positive CAP input jack.
3. Touch the test leads to the capacitor being tested.
4. Read the capacitance value on the LCD display.

    The meter will automatically adjust the ranges

    between nF, µF, and mF. It may take up to a

    minute to get a stable reading on large capacitors. 
5. If the symbol dISC appears on the LCD display,

    discontinue test and safely discharge capacitor

    before retesting. 

Capacitance Measurements

 

               

Safely discharge capacitors

before taking capacitance measurements. 

Summary of Contents for 12070T

Page 1: ...ciones de uso TESTING EQUIPMENT E361819 12 14 Rev 1 12070T manual Contents Made in China Product distributed by Southwire Company LLC One Southwire Drive Carrollton GA 30119 2014 Southwire Company LLC...

Page 2: ...attery compartment door firmly to maintain the waterproof and dust proof integrity of the meter Loose or overtightened screws or an improperly seated o ring may compromise the meter s water and dust i...

Page 3: ...ines to detached buildings Incoming service entrance from utility Outdoor pumps WARNING Operation is limited to CAT II applications when the insulated tips are removed from one or both test probes Ref...

Page 4: ...icon will be displayed on the LCD Momentarily press the HOLD button again to return to normal operation 6 Operation Backlight Button The backlight illuminates the display when the ambient light is too...

Page 5: ...rallel to the circuit under test For DC measurements touch the red test probe to the positive side of the circuit and the black test probe to the negative side of the circuit 5 Read the voltage on the...

Page 6: ...OL The resistance will be shown on the LCD display if it is below 400 10A 30 sec MAX CAT III 1000V CAT IV 600V FUSED MAX 400mA Temp Auto Power Off Waterproof every 15 min mA COM Hz CAP 12070T Autorang...

Page 7: ...s or an improperly seated o ring may compromise the meter s water and dust ingress protection WARNINGS To avoid electric shock do not operate your meter until the battery and fuse cover are in place a...

Page 8: ...mA 1 2 of rdg 5 digits 2 5 of rdg 5 digits Overload Protection FF500mA 1000V and F10A 1000V Fuse DC Current Auto ranging 400 0uA 4000uA 40 00mA 400 0mA 10A 0 1uA 1uA 10uA 100uA 10mA 1 0 of rdg 3 digit...

Page 9: ...tenance or repair operation outside of specifications or abnormal handling of the product Southwire s sole liability and the purchaser s exclusive remedy for any breach of this warranty is expressly l...

Page 10: ...AC rms 10A DC 1000VAC rms 1000VDC 1000VAC rms 1000VDC Lea entienda y siga todas las Reglas de Seguridad e Instrucciones de Operaci n en este manual antes de usar este metro Las caracter sticas de seg...

Page 11: ...binaci n del metro cables de pruebas y cualquier accesorio conectado al metro y cables de pruebas La combinaci n de clasificaci n es la M S BAJA de cualquier componente individual ADVERTENCIA El funci...

Page 12: ...neamente el bot n HOLD otra vez para regresar a la operaci n normal 6 Bot n de PEAK Hold Retenci n de picos La funci n Peak Hold captura el voltaje o corriente AC o DC m s alto Puede capturar alturas...

Page 13: ...positiva mA 4 Para medidas de corriente hasta 4000 A mueva el interruptor rotatorio a la posici n A e inserte la sonda roja en la entrada positiva A 5 Presione moment neamente el bot n MODO para sele...

Page 14: ...oximadamente 30 o menos sonar un tono Si el circuito est por encima de los 400 la pantalla indicar OL La resistencia aparecer en la pantalla si est por debajo de los 400 10 Prueba de Diodos Nunca comp...

Page 15: ...cubierta trasera y la puerta de compartimiento de la bater a para mantener la integridad de las caracter sticas de impermeabilidad y a prueba de polvo del metro Tornillos flojos o demasiado apretados...

Page 16: ...10uA 100uA 10mA 1 0 de la lectura 3 digitos 2 5 de la lectura 3 digitos Protecci n de Sobrecarga FF500mA 1000V y F10A Fusible 1000V Entrada M xima 4000uA DC en un rango uA 400mA DC en un rango mA 10A...

Page 17: ...uier incumplimiento de esta garant a est limitada expresamente a la reparaci n o reemplazo del producto por parte de Southwire La reparaci n o reemplazo del producto se har bajo la determinaci n de So...

Reviews: