Test frequency: 100Hz, 1kHz ,10kHz and 100kHz
Level: 1V
Range: AUTO
Bias: 0V
Speed: slow
Short correction and open correction should be done before the test. Connect the standard
resistor (10Ω, 100Ω, 1kΩ, 10kΩ, and 100kΩ) to the test fixture. Change the test frequency.
The error between the test result and the standard value must be in the range of the |Z|
accuracy referring to this chapter.
6.6.10The DCR accuracy
Test condition
:
Function: DCR
Test frequency: -----
Level: -----
Range: AUTO
Bias: -----
Speed: slow
Short correction and open correction should be done before the test. Connect the standard
resistor (0.1 Ω , 1Ω, 10Ω, 100Ω, 1kΩ, 10kΩ, and 100kΩ) to the test fixture. The error
between the test result and the standard value must be in the range of the DCR accuracy
referring to this chapter.