ST2826 Operation manual
Ver1.0
◇
118
e.g.
:
WrtCmd(―COMP:BIN:COUN ON‖)
query syntax
:
COMParator:BIN:COUNt[STATe]?
return
:
<NR1><NL^END>
:BIN:COUNt:DATA?
is used to query the BIN count result.
query BIN
:
COMParator:BIN:COUNt:DATA?
return
:
<BIN1 count>,<BIN2 count>,…,<BIN9 count>,<OUT OF BIN count>,
<AUX BIN count><NL^END>
<BIN1-9 count>:NR1 format, is the count result of BIN 1 to 9.
<OUT OF BIN count>: NR1 format
,
is the count result of over the standard.
<AUX BIN count>: NR1 format
,
is the count result of the auxiliary BIN.
:BIN COUNT:CLEar
is used to clear all the count result.
command syntax
:
COMParator:BIN:COUNt:CLEar
e.g.
:
WrtCmd(―COMP:BIN:COUN:CLE‖)
8.1.16Mass MEMory subsystem command set
Mass MEMory
subsystem command set is used to save and load the file.
Command hierarchical chart:
Mass MEMory
:LOAD
:STATe <record number>
:STORe
:STATe <record number>
:LOAD:STATe
is used to load the saved file.
command syntax
:
MMEMory:LOAD:STATe <value>
<value>:the file sequence number of 0 to 9 (NR1 format).
e.g.
:
WrtCmd(―MMEM:LOAD:STAT 1‖);
:STORe:STATe
is used to save the current setting for the instrument as one file.
command syntax
:
MMEMory:STOR:STATe <value>
<value>: the file sequence of 0 to 9 (NR1 format) .
e.g.
:
WrtCmd(―MMEM:STOR:STAT 1‖);
8.2The GPIB common command for ST2826
●*RST
●*TRG
●*IDN
●*TST
Summary of Contents for ST2826
Page 2: ......
Page 10: ......
Page 51: ...ST2826 Operation manual Ver1 0 41 Figure 3 15 ...
Page 61: ...ST2826 Operation manual Ver1 0 51 ...
Page 144: ...ST2826 Operation manual Ver1 0 134 Figure 9 8 the input circuit for control signal ...