ST2826 Operation manual
Ver1.0
◇
117
e.g.: WrtCmd(―COMP:ABIN ON‖)
query syntax
: COMParator:Auxiliary BIN
?
return
:
<NR1><NL^END>
:SWAP
is used to set the swap comparator mode ON/OFF of the main parameter and the
secondary parameter.
e.g.:function:Cp-D; SWAP mode: ON; the function will be changed to D-Cp.At the
same time, the limit setting on BIN 1 to 9 is the high limit and the low limit for D, the
setting of BIN 2 is the Cp limit. The meaning is that if ON is selected, the main
parameter and the secondary parameter will be compared by being interchanged; if
OFF is seleccted, the parameters will be compared by the primary order. The state of
the swap compatator mode can be queried.
command syntax
:
ON
COMParator:SWAP OFF
1
0
char ‗1‘(49) is equivalent to ON
char ‗0‘(48) is equivalent to OFF
e.g.
:
WrtCmd(―COMP:SWAP ON‖)
query syntax
:
COMParator:SWAP?
return
:
<NR1><NL^END>
:BIN:CLEar
is used to clear the setting limit of every BIN in the limit list.
command syntax
:
COMParator:BIN:CLEar
e.g.
:
WrtCmd(―COMP:BIN:CLE‖)
:BIN:COUNT[:STATe]
is used to set the BIN counter ON/OFF. The state of the BIN counter
can be queried.
command syntax
:
ON
COMParator:BIN:COUNt[:STATe] OFF
1
0
char ‗1‘(49) is equivalent to ON
char ‗0‘(48) is equivalent to OFF
Summary of Contents for ST2826
Page 2: ......
Page 10: ......
Page 51: ...ST2826 Operation manual Ver1 0 41 Figure 3 15 ...
Page 61: ...ST2826 Operation manual Ver1 0 51 ...
Page 144: ...ST2826 Operation manual Ver1 0 134 Figure 9 8 the input circuit for control signal ...