ST2826 Operation manual
Ver1.0
◇
91
Command syntax
:
ON
AMPLitude:ALC OFF
1
0
char ‗1‘(49) is
equivalent to ON
char ‗0‘(48) is equivalent to OFF.
e.g.
:
WrtCmd(―AMPL:ALC 0‖); set the ALC close
query syntax
:
AMPLitude:ALC
?
return
:
<NR1><NL^END>
8.1.6Output RESister subsystem command set
Output RESister
subsystem command set is used to set the output
internal resistance mode. Char
‗?‘can be used to query the output internal resistance state.
command syntax
:
30
ORESister
100
e.g.
:
WrtCmd(―ORES 30‖);set the
output internal resistance
to 30 OHM
query syntax
:
ORESister?
return
:
<NR1><NL^END>
8.1.7OUTPut subsystem command set
OUTPut
subsystem command set is used to set the DC 100mA/10V bias source, and the DC
isolation
.
Command hierarchical chart:
OUTPut
:High POWer ON(1)
OFF(0)
:DC
:ISOLation
ON
(
1
)
OFF
(
0
)
:HPOWer
is used to set the DC 100mA/10V bias ON/OFF. Char ‗?‘ cand be used to query the
state of the bias source.