![Sourcetronic ST2826 Operation Manual Download Page 114](http://html1.mh-extra.com/html/sourcetronic/st2826/st2826_operation-manual_1336370114.webp)
ST2826 Operation manual
Ver1.0
◇
104
<value>: NR1,NR2 or NR3 format
,
0 to 60 s with 1ms as the resolution.
MIN : set the delay time to 0 s.
MAX:set the delay time to 60 s.
e.g.
:
WrtCmd(―TRIG:DEL 5S‖); set the delay time to 5 s.
query syntax
:
TRIGger:DELay?
return
:
<NR3><NL^END>
8.1.13FETCh? subsystem command set
FETCh?
subsystem command set is used to transmit a test result.
Command hierarchical chart:
FETCh
[:IMP]?
:Source MONitor
:VAC?
:IAC?
:fres :acur?
:bcur?
[:IMP]?
is used to notify the ST2826 to transmit the test result to the buffer.
query syntax
:
FETCh[:IMP]?
e.g.
:
WrtCmd(―TRIG:SOUR BUS‖);
WrtCmd(―TRIG‖);
WrtCmd(―FETC?‖);
ST2826 transmits the result with ASCII, the more details are below:
SN.NNNNNESNN
,
SN.NNNNNESNN
,
SN
,
SN or SNN NL^END
<DATA A>
<DATA B>
<state> <BIN NO.>
Figure 8-1 the ASCII format 1
<DATA A>,<DATA B> format
:
<DATA A>
(
the main parameter
)
,
<DATA B>
(
the
secondary parameter
)
are ASCII code with 12 bits
:
SN.NNNNNESNN
(
S
:
+/-
,
N
:
0 to 9
,
E
:
Exponent Sign
)
<state>format: when the following value is selected
,
<state> will display the test state.
Summary of Contents for ST2826
Page 2: ......
Page 10: ......
Page 51: ...ST2826 Operation manual Ver1 0 41 Figure 3 15 ...
Page 61: ...ST2826 Operation manual Ver1 0 51 ...
Page 144: ...ST2826 Operation manual Ver1 0 134 Figure 9 8 the input circuit for control signal ...