59
field will display A, the cursor will be moved to the LOW field
automatically.
j)
Press [3][2][5], +325 will be displayed in the bottom line of the screen.
The usable units (p, n, u, m and MORE 1/2) will be displayed in the
softkey area. Press n, the value of the field will be changed to 325.000n.
The cursor will be moved to the HIGH field of the sweep point 1
automatically.
k)
Press [3][3][3], +333 will be displayed in the bottom line of the screen.
The usable units (p, n, u, m and MORE 1/2) will be displayed. Press n, the
value of this field will be changed to 333.000n. The cursor will be moved
to the parameter area of the sweep point 2.
l)
Press [10], +10 will be displayed in the bottom line of the screen. The
usable units (Hz, kHz and MHz) will be displayed in the softkey area.
Press kHz, the value of this field will be changed to 10.0000k.
m)
Press [ ]. Move the cursor to the LMT field of the sweep point 2.
Current field displays , the softkey area will display LIMIT DATA A,
LIMIT DATA B and OFF.
n)
Press LIMIT DATA B to select the secondary compared parameter D. The
field will display , the cursor will be moved to the LOW field of the
sweep point 2 automatically.
o)
Press [0][.][0][0][0][1], +0.0001 will be displayed in the bottom line of the
screen. The usable units (p, n, μ, m and MORE 1/2) will be displayed in
the softkey area. Press MORE 1/2. k, M, *1 and MORE 2/2 will be
displayed. Press *1, the value of this field will be changed to 100.000μ.
The cursor will be moved to the HIGH field of the sweep point 2
automatically.
p)
Press [0][.][0][0][0][3], +0.0003 will be displayed in the bottom line of the
screen, the usable units (k, M, *1 and MORE 2/2) will be displayed in the
softkey area. Press *1, the value of the field will be changed to 300.000μ.
The cursor will be moved to the parameter area of the sweep point 3.
q)
According to steps l to p, Input the parameters of sweep point 3 in turn:
100 kHz, B, 0.0060 and 0.0100.
3) Setting for alarm
(a)Press [SYSTEM]. The SYSTEM SETUP page will be displayed.
(b)Move the cursor to beep field. The current field displays LOW. OFF, LOW and HIGH
will be displayed in the softkey area.
(c)Press HIGH to select beep with high sound