![Sourcetronic ST2826 Operation Manual Download Page 134](http://html1.mh-extra.com/html/sourcetronic/st2826/st2826_operation-manual_1336370134.webp)
ST2826 Operation manual
Ver1.0
◇
124
is less than the low limit in BIN1 to BIN9.
21
/SREJ
The secondary parameter is not good: the
test result is not in the range of the high
limit and the low limit.
22
23
24
NC
NC
NC
Not connect.
25
/KEY LOCK
When this single is effective, the keys in
the front panel are locked.
27
28
EXT.DCV1
The external DC voltage 1
:
the pull-up
DC power provider pin for optoelectronic
coupling signal(/BIN-/BIN9,/AUX
,
/OUT
,
/PHI
,
/PLO
,
/SREJ).
29
/ALARM
When circuit is interrupted, /ALARM is
effective.
30
/INDEX
When the analog test is finished and the
UNKNOWN terminal can be connected to
another DUT, /INDEX is effective. But the
comparison signal is effective until /EOM
is effective.
31
/EOM
End Of Measurement
:
when the test data
and the compared result are effective, this
signal is effective.
32,33
COM2
The reference ground for external power
EXTV2.
34,35,36
COM1
The reference ground for external power
EXTV1.