Service Manual: FUSION EQUINE DR® |
43
2. Safety, Warranty, and Licensing Information
Ensure that all mounting and fastening hardware is tightened properly, and that all securing
mechanisms on connectors and covers are properly latched.
Inspect all cabling, mounting, and securing mechanisms during each Preventive Maintenance
(PM) cycle to ensure that electrical connections and other hardware do not become loose over
time.
Some components of the digital radiography system are of significant size and weight. Observe
appropriate lifting and handling techniques when moving heavy equipment or components.
Obtain assistance when necessary to avoid injury to persons or damage to equipment.
Electrical Safety
Electrical power sufficient to cause injury or death is present inside many of the digital
radiography system components whenever they are connected to AC power. Take appropriate
safety precautions, use safety disconnects (such as fuses or breakers) wherever possible, and
disconnect AC supply cables from components prior to removing covers for maintenance or
service.
Caution:
Internal power supplies contain capacitors that may remain
charged for a period of time after the power source is removed.
Before performing work inside any of the enclosures of digital
radiography system components, wait at least 60 seconds after
removing the AC supply cable for complete discharge.
Alimentations internes contiennent des condensateurs qui peuvent
rester chargés pour une période de temps après que la source
d'alimentation est débranché. Avant d'effectuer tout travail à l'intérieur
des enceintes de composants du système x -ray, attendez au moins
60 secondes après avoir retiré le câble d'alimentation CA pour une
décharge complète.
Components inside enclosures of the digital radiography system are sensitive to
electrostatic discharge (ESD). Personnel servicing components of the digital radiography
system must take appropriate ESD prevention measures to minimize the risk of damage to
system hardware.
Do not block or restrict airflow into or out of the computer or the enclosure around the detector,
if applicable. Adequate air cooling is required to prevent overheating the components inside
these enclosures.
Apply measures to prevent liquids, particularly toxic or hazardous fluids, from coming into
contact with the digital radiography system components and equipment. When cleaning the
digital radiography system equipment, do not spray or pour fluid directly onto equipment
surfaces. Use a soft cloth, dampened lightly with a cleaning solution, and gently wipe
system components.
When electrical components must be replaced, use only components that are appropriately
rated for the application. Replace fuses, switches, or connectors only with components of the
same type and rating as the original equipment.
To avoid electric shock, the digital radiography system must be powered from an AC supply
circuit that includes an adequate earth ground. Connect the digital radiography system
components only to receptacles labeled or marked as medical grade.
Warning:
The digital radiography system and its components are
designed to be connected to a properly grounded AC supply sufficient
to support system operation. Using power strips or other multiple-
socket outlets
Summary of Contents for FUSION Equine DR II
Page 2: ... 2022 Sound company All rights reserved This page left blank intentionally ...
Page 10: ......
Page 35: ...Service Manual FUSION EQUINE DR II 17 1 System Overview ...
Page 38: ...20 Service Manual FUSION EQUINE DR II 1 System Overview Figure 10 Ready Indicator dimensions ...
Page 75: ...Service Manual FUSION EQUINE DR II 57 3 Installing the FUSION EQUINE DR II System ...
Page 77: ...Service Manual FUSION EQUINE DR II 59 3 Installing the FUSION EQUINE DR II System ...