144 | Service Manual: FUSION EQUINE
DR®
5. Maintaining the FUSION EQUINE DR
®
X-ray
System
Cleaning the Digital Radiography System
About this task
The digital radiography system is designed and suitable for use in typical clinical environments.
During use, the system, all peripherals, and the detector should be adequately protected against
spilled or splashed fluids and should therefore not require disinfection beyond routine cleaning
as part of preventive maintenance of the equipment.
Cleaning performed during preventative maintenance requires only compressed air and a
mild soap and water solution. If disinfection is desired or becomes necessary, a disinfectant
solution may be used in place of soap and water to clean the digital radiography system
equipment. In either case, prepare the solution in accordance with instructions provided by the
manufacturer of the cleaning agent.
Warning:
Do not pour or spray liquid directly onto any component of
the digital radiography system. Apply the cleaning agent to a clean
cloth and gently wipe the equipment to clean.
Warning:
Ne pas verser ou vaporiser de liquide directement sur l'un
des composants du système digital radiography. Appliquer l'agent de
nettoyage sur un chiffon propre et essuyez doucement l'équipement à
nettoyer.
Procedure
Review the information in the following topics, and perform cleaning and maintenance
tasks in accordance with the information provided. Cleaning and preventative maintenance
should be performed approximately every six months or as required by the site.
on page 144
on page 145
c.
Removing Dust From Fans and Heatsinks
on page 145
Approved Disinfection Agents
Any EPA-registered agent classified as a low- or intermediate-level product for hard, non-
porous surfaces and equipment may be used. Prepare and use disinfectants in accordance
with manufacturer’s instructions.
Summary of Contents for FUSION Equine DR II
Page 2: ... 2022 Sound company All rights reserved This page left blank intentionally ...
Page 10: ......
Page 35: ...Service Manual FUSION EQUINE DR II 17 1 System Overview ...
Page 38: ...20 Service Manual FUSION EQUINE DR II 1 System Overview Figure 10 Ready Indicator dimensions ...
Page 75: ...Service Manual FUSION EQUINE DR II 57 3 Installing the FUSION EQUINE DR II System ...
Page 77: ...Service Manual FUSION EQUINE DR II 59 3 Installing the FUSION EQUINE DR II System ...