Service Manual: FUSION EQUINE DR® |
139
5. Maintaining the FUSION EQUINE DR
®
X-ray
System
8.
The software will display the Dosage % and determine if the exposure was a success of failure.
Note:
Failure of a successful calibration exposure consists on being too high or
too low. Adjust technique on the generator and retake the calibration shot.,
you have a 3 to 5 second window to take an exposure.
Successful Calibration Exposure
Failed Calibration Exposure