![Soredex DIGORA Optime DXR-50 001 User Manual Download Page 8](http://html1.mh-extra.com/html/soredex/digora-optime-dxr-50-001/digora-optime-dxr-50-001_user-manual_1332635008.webp)
2
User’s manual 204253
Imaging plate unit
Introduction
NOTES:
Only personnel trained and authorized by the
manufacturer of the unit are allowed to install and
configure the unit.
Only use the imaging plates, protective covers and
hygiene bags supplied by the manufacturer of the unit.
Please read the section
7. Warnings and
precations
before using the unit.
1.2 System installation
Positioning the unit
Do not position the unit in direct sunlight or near bright
light. Sunlight or bright light must not be allowed to
shine directly on the unit door into which the IPs are
inserted.
Position the unit on a stable flat surface so that
vibrations will not degrade the image quality.
The unit can also be attached to a wall, under or on a
shelf with the optional mounting kit.
The unit must not be positioned so that it it touching
other equipment. It must not be placed on top of or
under other equipment.
The unit can be positioned within the environment in
which the patient is examined and treated (patient
environment).