2. Intraoral imaging plate unit
Imaging plate unit
Retrieve last image
If the last image read is not transferred to the PC
because of a network, communication, PC or
software failure, the last image read can be retrieved.
IMPORTANT NOTE
The LAST read image can only be retrieved if the unit
is left on.
If the unit is switched off the image will
be lost
.
To retrieve the last scanned image:
1. Correct the problem that caused the
communication failure. When the connection
between the unit and the PC is re-established the
last read image will automatically be transferred
to the PC.
2.
PC:
If the image is not automatically transferred to
the PC, select the
Setup
>
Scanner
page from
the dental imaging software your are using.
3.
PC:
In the
Retrieve Last Image
field, click the
check box to retrieve the last read image.
NOTE:
If required you can select different parameters
(e.g. resolution, show image preview etc.) for the
image to be retrieved.
4.
PC:
Click
OK
to close the
Setup
window.
The last read image will be transferred to the PC.