28
User’s manual 204253
Imaging plate unit
4. Handling and care of imaging plates
4.4 Storage
•
Unpacked, exposed to ambient light in the
dedicated storage box
•
Below 33°C / 80% RH and shielded from X-rays
and ultraviolet radiation.
•
If an IP is stored for over 24 hours in a hygiene
bag or in a location that is shielded from ambient
light, the IP must be erased, to remove any
potential fogging, before being used to take an
exposure.
4.5 Replacement
Replace an IP if:
•
the image shows marks, spots or dots, which still
appear on images even after the IP has been
properly cleaned.
•
the IP is mechanically damaged (scratched) or
badly bent.
4.5 Disposal
IPs must be disposed of in accordance with all local,
national and international regulations regarding the
disposal of non-environmentally friendly or hazardous
materials.
Phosphor substance, under the top coat, on the
sensitive side of IPs must not be swallowed.