User’s manual 204253
A - 1
A. Technical Specifications
Imaging plate unit
Appendix A. Technical Specifications
A.1 Unit
Intraoral imaging plate unit
DXR 50 Classification
IEC60601-1
- Class 1 or 2 equipment depending on the classification
of the PSU. No applied part
- Continuous operation
- IPX0 (enclosed equipment without protection against
ingress of liquids
Laser Safety Classification
CLASS 1 LASER PRODUCT
EN 60825-1 :2007
Dimensions (H x W x D)
190 mm x 200 mm x 383mm (7.5in x 7.8in x 15.1in)
Weight
9.8 – 10.3 kg (21.6 – 22.7 lb)
Power supply unit (PSU)
POWERBOX EMX 805121
Operating voltage
24 VDC (PSU: 100 – 240 VAC, 50/60 Hz)
Operating current
Less than 1.5 A
Pixel size, selectable
35
µ
m (Super), 64
µ
m (High)
Bit depth
14 bits grayscale
Clinical resolution
10 lp/mm
Spatial resolution
14.3 lp/mm
Interface cable
For Ethernet connections, use an unshielded CAT6
LAN cable, so that multiple chassis must not be
connected! The PC / Ethernet switch to which unit is
connected to should be approved appropriately (e.g.
EN 60950, IEC 60950, UL 60950).
Operating environment
+10
°
C - +40
°
C, 30 – 90 RH%, 700 – 1060 mbar
Storage / transportation
environm ent
-10
°
C – +50
°
C, 0 – 90 RH%, 500 – 1080 mbar