CRANEX
®
Novus e
8. Aligning the unit
Service
manual
109
8.4 Checking and adjusting the Chin Support
The Chin Support is in the correct positioned when
the beam strikes the middle line on the alignment
plate when the rotating unit is parallel to the
side
of
the upper shelf.
1. Make sure that the you are able able to see
the beam alignment tool from beneath the
tubehead.
2. Make sure that
exp
(exposure, no rotation,
CCD sensor not active) function has been se-
lected (
CRANEX
®
Novus e service terminal
).
3. Place the alignment plate into the two holes at
the front of the ball pin phantom.
4. Manually rotate the Rotating Unit so that it is
parallel to the
side
of the upper shelf and the
tubehead is at the front. Insert a 4 mm hexa-
gon socket wrench into the locking hole in the
Rotation bearing to hold the rotating unit in
position.
Summary of Contents for Cranex Novus e
Page 1: ...Service Manual for the CRANEX Novus e Dental X ray Unit 208710 rev 1 2013 01...
Page 2: ......
Page 23: ...14 Service Manual 2 Unit Description CRANEX Novus e 2 3 Unit dimensions...
Page 34: ...CRANEX Novus e 2 Unit Description Service manual 25 2 7 Block diagram...
Page 35: ...26 Service manual 2 Unit Description CRANEX Novus e 2 8 Fuse diagram...
Page 40: ...CRANEX Novus e 3 Circuit Boards L1200 CPU Board Service manual 31 L1200 Block Diagram...
Page 43: ...34 Service manual 3 Circuit Boards L1500 CCD Power Supply CRANEX Novus e L1500 Block diagram...
Page 47: ...38 Service manual 3 Circuit Board L1600 Z Motor Driver CRANEX Novus e L1600 Block diagram...
Page 51: ...42 Service manual 3 Circuit Boards L1700 Connector Board CRANEX Novus e L1700 Block Diagram...
Page 54: ...CRANEX Novus e 3 Circuit Boards L1800 Generator Board Service manual 45 L1800 Block diagram...
Page 58: ...CRANEX Novus e 3 Circuit Boards P1910 PFC Board Service manual 49 P1910 Block diagram...