Error E19 (Exposure switch stuck down)
CAUSE
Exposure switch stuck down during unit start up.
SOLUTION
Switch the unit off and check that the exposure switch
is not stuck in the exposure position. Switch the unit on
again. If the message reappears, the switch is faulty,
replace.
Error E20 (PCF: POWERGOOD signal is down)
CAUSE
No power signal.
SOLUTION
Check to see if the signal is present on L1500.
If the signal is not present, replace P1910.
If signal is present, the fault could be in the signal rout-
ing. Check signal on L1500, L1700 and L1200 and in
the relevant cables. Replace any faulty boards or ca-
bles.
Error E22 (Button stuck)
Appears only on start-up
CAUSE
The membrane switches on the column are stuck.
SOLUTION
Reboot, if probles persists, replace the membrane but-
tons.
Summary of Contents for Cranex Novus e
Page 1: ...Service Manual for the CRANEX Novus e Dental X ray Unit 208710 rev 1 2013 01...
Page 2: ......
Page 23: ...14 Service Manual 2 Unit Description CRANEX Novus e 2 3 Unit dimensions...
Page 34: ...CRANEX Novus e 2 Unit Description Service manual 25 2 7 Block diagram...
Page 35: ...26 Service manual 2 Unit Description CRANEX Novus e 2 8 Fuse diagram...
Page 40: ...CRANEX Novus e 3 Circuit Boards L1200 CPU Board Service manual 31 L1200 Block Diagram...
Page 43: ...34 Service manual 3 Circuit Boards L1500 CCD Power Supply CRANEX Novus e L1500 Block diagram...
Page 47: ...38 Service manual 3 Circuit Board L1600 Z Motor Driver CRANEX Novus e L1600 Block diagram...
Page 51: ...42 Service manual 3 Circuit Boards L1700 Connector Board CRANEX Novus e L1700 Block Diagram...
Page 54: ...CRANEX Novus e 3 Circuit Boards L1800 Generator Board Service manual 45 L1800 Block diagram...
Page 58: ...CRANEX Novus e 3 Circuit Boards P1910 PFC Board Service manual 49 P1910 Block diagram...