3. Circuit Boards - L1800, Generator Board
CRANEX
®
Novus e
L1800 - Description
The Generator board receives kV and mA reference
signals from L1200 as PWM modulated signal. Based
on the kV-reference value, the Generator board gener-
ates the corresponding high voltage between the cath-
ode and anode of the x-ray tube. Based on milliampere
reference value, the generator board generates pre-
heat current (to warm up the
fi
lament before x-rays are
switched on) and
fi
lament current (during exposure).
The Generator board receives kV and mA feedback
signals from the tubehead that are used to monitor and
adjust the generated values. The Generator board pro-
duces its own supply voltages.
Summary of Contents for Cranex Novus e
Page 1: ...Service Manual for the CRANEX Novus e Dental X ray Unit 208710 rev 1 2013 01...
Page 2: ......
Page 23: ...14 Service Manual 2 Unit Description CRANEX Novus e 2 3 Unit dimensions...
Page 34: ...CRANEX Novus e 2 Unit Description Service manual 25 2 7 Block diagram...
Page 35: ...26 Service manual 2 Unit Description CRANEX Novus e 2 8 Fuse diagram...
Page 40: ...CRANEX Novus e 3 Circuit Boards L1200 CPU Board Service manual 31 L1200 Block Diagram...
Page 43: ...34 Service manual 3 Circuit Boards L1500 CCD Power Supply CRANEX Novus e L1500 Block diagram...
Page 47: ...38 Service manual 3 Circuit Board L1600 Z Motor Driver CRANEX Novus e L1600 Block diagram...
Page 51: ...42 Service manual 3 Circuit Boards L1700 Connector Board CRANEX Novus e L1700 Block Diagram...
Page 54: ...CRANEX Novus e 3 Circuit Boards L1800 Generator Board Service manual 45 L1800 Block diagram...
Page 58: ...CRANEX Novus e 3 Circuit Boards P1910 PFC Board Service manual 49 P1910 Block diagram...