90
Service
manual
5. S2terminal and service commands
CRANEX
®
Novus e
stillmode
still image exposure
Generates x-rays and activates the CCD sensor but the
rotating unit does not move.
It is used to check the beam alignment during unit
installation (refer to the Installation manual for more
information).
CAUTION:
Protect yourself from radiation when carrying out
this procedure.
Press and hold the exposure switch to take a still image
exposure.
NOTE:
The "
exp
" command, described earlier, generates x-
rays but does not activate the sensor or move the rotat-
ing unit.
warmup
tubehead warmup before fi rst use
Initializes the tubehead. Must be used when the unit is
installed or if the unit has not been used for a long time.
CAUTION:
Protect yourself from radiation when carrying out
this procedure.
Press and hold the exposure switch until the procedure
is complete. The procedure may take some minutes.
Summary of Contents for Cranex Novus e
Page 1: ...Service Manual for the CRANEX Novus e Dental X ray Unit 208710 rev 1 2013 01...
Page 2: ......
Page 23: ...14 Service Manual 2 Unit Description CRANEX Novus e 2 3 Unit dimensions...
Page 34: ...CRANEX Novus e 2 Unit Description Service manual 25 2 7 Block diagram...
Page 35: ...26 Service manual 2 Unit Description CRANEX Novus e 2 8 Fuse diagram...
Page 40: ...CRANEX Novus e 3 Circuit Boards L1200 CPU Board Service manual 31 L1200 Block Diagram...
Page 43: ...34 Service manual 3 Circuit Boards L1500 CCD Power Supply CRANEX Novus e L1500 Block diagram...
Page 47: ...38 Service manual 3 Circuit Board L1600 Z Motor Driver CRANEX Novus e L1600 Block diagram...
Page 51: ...42 Service manual 3 Circuit Boards L1700 Connector Board CRANEX Novus e L1700 Block Diagram...
Page 54: ...CRANEX Novus e 3 Circuit Boards L1800 Generator Board Service manual 45 L1800 Block diagram...
Page 58: ...CRANEX Novus e 3 Circuit Boards P1910 PFC Board Service manual 49 P1910 Block diagram...