![Soredex Cranex Novus e Service Manual Download Page 75](http://html1.mh-extra.com/html/soredex/cranex-novus-e/cranex-novus-e_service-manual_1332638075.webp)
66
Service
manual
4. Troubleshooting
CRANEX
®
Novus e
Error E8 (File system failure)
CAUSE
GUI graphics are corrupted.
SOLUTION
Update the GUI graphics.
Error E16 (Z lift not allowed due to duty cycle)
CAUSE
Z-motor over heated.
SOLUTION
Allow motor to cool down.
Error E18 (Display not connected)
Check Service Assistant for information about this
error code
CAUSE A
Cable from L1500 to L900 faulty.
SOLUTION A
Replace cable.
CAUSE B
L1500 faulty.
SOLUTION B
Replace L1500.
CAUSE C
Cable between L1200 and L900 faulty.
SOLUTION C
Replace cable.
CAUSE D
Cable between L900 and LCD display faulty.
SOLUTION D
Replace cable.
CAUSE E
L900 faulty.
SOLUTION E
Replace L900.
CAUSE F
LCD display is faulty.
SOLUTION E
Replace LCD display.
Summary of Contents for Cranex Novus e
Page 1: ...Service Manual for the CRANEX Novus e Dental X ray Unit 208710 rev 1 2013 01...
Page 2: ......
Page 23: ...14 Service Manual 2 Unit Description CRANEX Novus e 2 3 Unit dimensions...
Page 34: ...CRANEX Novus e 2 Unit Description Service manual 25 2 7 Block diagram...
Page 35: ...26 Service manual 2 Unit Description CRANEX Novus e 2 8 Fuse diagram...
Page 40: ...CRANEX Novus e 3 Circuit Boards L1200 CPU Board Service manual 31 L1200 Block Diagram...
Page 43: ...34 Service manual 3 Circuit Boards L1500 CCD Power Supply CRANEX Novus e L1500 Block diagram...
Page 47: ...38 Service manual 3 Circuit Board L1600 Z Motor Driver CRANEX Novus e L1600 Block diagram...
Page 51: ...42 Service manual 3 Circuit Boards L1700 Connector Board CRANEX Novus e L1700 Block Diagram...
Page 54: ...CRANEX Novus e 3 Circuit Boards L1800 Generator Board Service manual 45 L1800 Block diagram...
Page 58: ...CRANEX Novus e 3 Circuit Boards P1910 PFC Board Service manual 49 P1910 Block diagram...