CRANEX
®
Novus e
3. Circuit Boards - L1600, Z-Motor Driver
Service
manual
39
L1600 - Indicator lamp
Lamp
Function
Indicates
LA1 GLIM
AC indicator lamp L1600 receiving line voltage.
L1600 - Indicator LEDs
LED
Colour Indicates
D1
green
+5V
on
D2
green
ZACT-movement
key
pressed
D3
green
ZON
on
D4
red
ESTOP on. Emergency stop button is on.
D5
green
SET 100V: autotransformer tapping enabled
D6
green
SET 115V: autotransformer tapping enabled
D7
green
SET 230V: autotransformer tapping enabled
L1600 - Test Points
Number Signal Value
TP1
ESTOP_5V
+5V normally, 0 ... +0.5V when emergency
button
pressed
down
TP2
ZDIR
0V driving up, +4.8V...+5V driving down
TP3
ZON
+3V when up/down key pressed, 0V when not.
TP4
+5V
+5V
TP5
GND
0V
TP6
ZACT*
0 ... +0.5 when up/down key pressed,
otherwise
close
to
+5V.
Summary of Contents for Cranex Novus e
Page 1: ...Service Manual for the CRANEX Novus e Dental X ray Unit 208710 rev 1 2013 01...
Page 2: ......
Page 23: ...14 Service Manual 2 Unit Description CRANEX Novus e 2 3 Unit dimensions...
Page 34: ...CRANEX Novus e 2 Unit Description Service manual 25 2 7 Block diagram...
Page 35: ...26 Service manual 2 Unit Description CRANEX Novus e 2 8 Fuse diagram...
Page 40: ...CRANEX Novus e 3 Circuit Boards L1200 CPU Board Service manual 31 L1200 Block Diagram...
Page 43: ...34 Service manual 3 Circuit Boards L1500 CCD Power Supply CRANEX Novus e L1500 Block diagram...
Page 47: ...38 Service manual 3 Circuit Board L1600 Z Motor Driver CRANEX Novus e L1600 Block diagram...
Page 51: ...42 Service manual 3 Circuit Boards L1700 Connector Board CRANEX Novus e L1700 Block Diagram...
Page 54: ...CRANEX Novus e 3 Circuit Boards L1800 Generator Board Service manual 45 L1800 Block diagram...
Page 58: ...CRANEX Novus e 3 Circuit Boards P1910 PFC Board Service manual 49 P1910 Block diagram...