8. Aligning the unit
Service
manual
107
6.
Protect yourself from radiation.
Darken the room and then position yourself
behind and below the tubehead.
Press and hold down the exposure switch.
Make a note of where the beam strikes the
alignment plate.
7. Remove the hexagon socket wrench from
the locking hole and then manually rotate the
Rotating Unit 180º so that it is parallel to the
front
of the upper shelf and the tubehead is on
the
right
. Reinsert the hexagon socket wrench
into the locking hole in the Rotation bearing to
hold the rotating unit in position.
8.
Protect yourself from radiation.
Press and hold down the exposure switch.
Make a note of where the beam strikes the
alignment plate on this side.
Summary of Contents for Cranex Novus e
Page 1: ...Service Manual for the CRANEX Novus e Dental X ray Unit 208710 rev 1 2013 01...
Page 2: ......
Page 23: ...14 Service Manual 2 Unit Description CRANEX Novus e 2 3 Unit dimensions...
Page 34: ...CRANEX Novus e 2 Unit Description Service manual 25 2 7 Block diagram...
Page 35: ...26 Service manual 2 Unit Description CRANEX Novus e 2 8 Fuse diagram...
Page 40: ...CRANEX Novus e 3 Circuit Boards L1200 CPU Board Service manual 31 L1200 Block Diagram...
Page 43: ...34 Service manual 3 Circuit Boards L1500 CCD Power Supply CRANEX Novus e L1500 Block diagram...
Page 47: ...38 Service manual 3 Circuit Board L1600 Z Motor Driver CRANEX Novus e L1600 Block diagram...
Page 51: ...42 Service manual 3 Circuit Boards L1700 Connector Board CRANEX Novus e L1700 Block Diagram...
Page 54: ...CRANEX Novus e 3 Circuit Boards L1800 Generator Board Service manual 45 L1800 Block diagram...
Page 58: ...CRANEX Novus e 3 Circuit Boards P1910 PFC Board Service manual 49 P1910 Block diagram...