33
Continued
Main
Code
Sub
Code
ERROR/ALARM indicator
Description
Error remedy method
Refer to section
Lighting
interval
Switching off
condition
02
120
1 second
Disc/Power
Optical block assembly
1Ach is recording data
with the maximum laser
output.
1. Eject the cartridge.
2. Check laser parameter 1 of hours meter.
When laser parameter 1 of hours meter has:
•
not reached the end of replacement cycle:
Go to step 3 in the error remedy method.
•
reached the end of replacement cycle:
It is likely laser in optical block assembly (OP-1) reached to the end of
its life cycle.
Replace optical block assembly (OP-1).
3. Startup the maintenance menu and perform [CHECK] > [LENS
CLEANING].
After [LENS CLEANING] is performed:
,
•
when the symptom is improved,
The problem is fixed. Use the unit as it is.
•
when the same symptom keeps occurring:
There is something wrong with the optical or electrical system in
optical block assembly (OP-1).
Replace optical block assembly (OP-1).
“2-1-2. Periodic Inspection and
125
0.25 seconds
Power
Optical block assembly
1Ach detected a laser
output error.
1. Eject the cartridge.
2. Check laser parameter 1 of hours meter.
When laser parameter 1 of hours meter has:
•
not reached the end of replacement cycle:
There is something wrong with the optical or electrical system in
optical block assembly (OP-1).
•
reached the end of replacement cycle:
It is likely laser in optical block assembly (OP-1) reached to the end of
its life cycle.
3. Replace optical block assembly (OP-1).
“2-1-2. Periodic Inspection and
127
0.25 seconds
Power
Laser drive overcurrent
was detected in optical
block assembly 1Ach.
Summary of Contents for PDW-U4
Page 7: ...4 Frame Wiring 151 Revision History 152 ...
Page 24: ...21 Hold the shaded portions Keep away a screwdriver Objective lens Actuator ...
Page 155: ...152 Revision History Date History Contents 2020 10 1st Edition 9 932 764 01 ...
Page 156: ...PDW U4 SY PDW U4 CN J E 9 932 764 01 Sony Corporation Printed in Japan 2020 10 08 2020 ...