32
Continued
Main
Code
Sub
Code
ERROR/ALARM indicator
Description
Error remedy method
Refer to section
Lighting
interval
Switching off
condition
02
0A5
0.25 seconds
Power
Optical block assembly
0Bch detected a laser
output error.
1. Eject the cartridge.
2. Check laser parameter 0 of hours meter.
When laser parameter 0 of hours meter has:
•
not reached the end of replacement cycle:
There is something wrong with the optical or electrical system in
optical block assembly (OP-0).
•
reached the end of replacement cycle:
It is likely laser in optical block assembly (OP-0) reached to the end of
its life cycle.
3. Replace optical block assembly (OP-0).
“2-1-2. Periodic Inspection and
0A7
0.25 seconds
Power
Laser drive overcurrent
was detected in optical
block assembly 0Bch.
0A8
0.25 seconds
Disc/Power
Optical block assembly
0Bch determines that
there is no cartridge, and
then the laser output is
stopped.
1. Eject the cartridge.
2. Startup the maintenance menu and display [CHECK] > [LOADER].
3. Insert the cartridge and check [DOWN SENSOR] and [SENSOR
HOLE].
•
when [DOWN SENSOR] is set to [DOWN] and [SENSOR HOLE]
displays “4”:
The problem is fixed. Use the unit as it is.
•
when [DOWN SENSOR] is set to other than [DOWN] and [SENSOR
HOLE] does not display “4”:
There is a possibility that a detection sensor is damaged.
Go to step 4 in the error remedy method.
4. Check the corresponding detection sensors, and then replace the
SE-1004A board or the SE-1005A board.
Summary of Contents for PDW-U4
Page 7: ...4 Frame Wiring 151 Revision History 152 ...
Page 24: ...21 Hold the shaded portions Keep away a screwdriver Objective lens Actuator ...
Page 155: ...152 Revision History Date History Contents 2020 10 1st Edition 9 932 764 01 ...
Page 156: ...PDW U4 SY PDW U4 CN J E 9 932 764 01 Sony Corporation Printed in Japan 2020 10 08 2020 ...