OPERATING MANUAL PAT-806 version 1.08
62
Power measurement S
Display range
Resolution
Measurement uncertainty
(basic)
0 VA...999 VA
1 VA
(5 % m.v. + 3 digits)
1 kVA
…3.99 kVA
0.01 kVA
adjustable measuring time in the range of: Cont , 1 s
…60 s with resolution of 1 s
Current consumption measurement
Display range
Resolution
Measurement uncertainty
(basic)
0.00 A ... 15.99 A
0.01 A
(2 % m.v. + 3 digits)
adjustable measuring time in the range of: Cont , 1 s
…60 s with resolution of 1 s
Measurement of voltage at the test socket
Display range
Resolution
Measurement uncertainty
(basic)
187.0 V ... 265.0 V
0.1 V
(2 % m.v. + 2 digits)
Other technical specification
a) type of insulation ...................................................... double, EN 61010-1 and IEC 61557 compliant
NOTE!
During the measurement of S, I
∆
, I
PE
and I
T
PE of the power supply socket is connect-
ed to PE of the test socket.
b) measurement category ........................................................................ II 300 V acc. to EN 61010-1
c) protection class of enclosure acc. to EN 60529 ....................................................................... IP40
d) power supply for the meter .............................................................................. 187 V
…265 V, 50 Hz
e) load current .......................................................................................................... max 16 A (230 V)
f) dimensions ........................................................................................ 330 mm x 235 mm x 120 mm
g) meter weight ............................................................................................................. approx. 5,0 kg
h) storage temperature................................................................................................
–20
C...+70
C
i)
operating temperature ................................................................................................. 0
C...+40
C
j)
humidity ....................................................................................................................... 20 %
…80 %
k) nominal temperature ............................................................................................... +20
C...+25
C
l)
reference humidity ....................................................................................................... 40 %
…60 %
m) altitude (above sea level) .................................................................................................. < 2000 m
n) display ...................................................................................................................... LCD, segment
o) memory of measurement results ....................................................................................... 990 cells
p) data transmission................................................................................................................ USB 2.0
q) quality standard .......................... development, design and manufacturing are ISO 9001 compliant
r) the product meets the EMC requirements acc. to ......... EN 61326-1:2009 and EN 61326-2-2:2006