OPERATING MANUAL PAT-806 version 1.08
50
Now you can choose performing of individual measurements for a given test and class.
Use
and
to choose performing the
measurement (
) or cancel it (
).
To change the parameters of the active
measurement press
SET
. Set the next pa-
rameters as in section 4.2. Confirm by press-
ing
ENTER
. Use
to activate and set pa-
rameters of the next measurement.
In this way, you may program the test by activating measurements in the following sequence: an
initial measurement → R
E
→ R
ISO
(1 - single, 3 - triple for welding machines)
→ I
SUB
→ I
PE
→ I
∆
→ I
T
→ I
L
(for welding machines
) → Ur (for welding machines) → Up (for welding machines) → S.
After approving the last change by pressing
ENTER
press it two more times and the
meter will enter standby mode displaying
the following screen.
Press
START
to start the measurement.
Notes:
- Readiness for next measurement is obtained automatically, but the individual measurements should
be initiated by pressing
START
, as in single measurements.
- The duration of the measurement may be reduced by pressing
STOP/ESC
. The result remains as at
the time of stopping the measurement and the meter turns into standby mode until the next meas-
urement. Double pressing of
STOP/ESC
interrupts the cycle of automatic measurement and all the
previous results are lost.