OPERATING MANUAL PAT-806 version 1.08
21
Symbol
indicates the need for visual inspection on the tested device. Please check the
power cord insulation, quality of housing and mains plug (for cracks or damages) etc.
Press
ENTER
, if this test is success-
ful (
is displayed) or
STOP/ESC
, when the test result is
negative (
is displayed).
Symbol
indicates the possibility to check the fuse removed from the tested device.
Contact the fuse with the test points. The efficiency of the fuse
is indicated by displaying
and an acoustic signal.
Notes:
- Tested device must be turned on.
- R
L-N
measurement is intended for resistance objects, in case of inductive objects, the result may be
burdened with an additional errors.
- Fuse test is possible when
READY
message is displayed.
- Do not touch the two metal ends of the fuse with your fingers during the test, because a blown fuse
may be diagnosed as good.
4.2 Measurement of protective lead resistance using 200 mA cur-
rent
Keep pressing
R
PE
button, until the screen in-
forming about the readiness for measurement
with 200 mA current is displayed.
When parameters must be changed press
SET
.