OPERATING MANUAL PAT-806 version 1.08
19
Window for configuring Auto-measurements.
The software enables user to read the current configuration of the meter, save the meter settings
into a file, upload the configuration from a file, create files with different configurations, which is a
simple way to prepare a number of configurations for various requirements, e.g. for different custom-
ers and quickly reset the meter according to current needs.
4 Measurements
Notes:
For the convenience of measurements on devices of Class II the socket marked with
symbol is connected with PE pin of the test socket. Do not connect dangerous vol-
tage to this socket.
"I
2
" banana socket is permanently connected to PE of IEC socket. Do not connect dan-
gerous voltage to this socket.
- Tested device must be turned on.
- Measurement activated with defined duration time =
- continuous measurement
– lasts as
long as
START
button is pressed. Maintaining the measurement is possible by pressing
ENTER
with
START
button kept pressed.
- Each measurement with duration =
- continuous measurement
– maintaining the measure-
ment by pressing
ENTER
may be stopped by pressing
STOP/ESC
.
- After completing each measurement use
,
to see parameters (limits) including the date and
time of measurement.
- All data can be entered using the program on your PC.