PAT-806-IT
– USER MANUAL
55
In this way, you may program the test by activating measurements in the following sequence: an
initial measurement → R
E
→ R
ISO
(1 - single, 3 - triple for welding machines)
→ I
SUB
→ I
PE
→ I
∆
→ I
T
→ I
L
(for welding machines
) → Ur (for welding machines) → Up (for welding machines) → S.
After approving the last change by pressing
ENTER
press it two more times and the
meter will enter standby mode displaying
the following screen.
Press
START
to start the measurement.
Notes:
- Readiness for next measurement is obtained automatically, but the individual measurements should
be initiated by pressing
START
, as in single measurements.
- The duration of the measurement may be reduced by pressing
STOP/ESC
. The result remains as at
the time of stopping the measurement and the meter turns into standby mode until the next meas-
urement. Double pressing of
STOP/ESC
interrupts the cycle of automatic measurement and all the
previous results are lost.
- If the result of one of the tests is improper (
), you may record it (finish the autotest by pressing
ENTER
and consider the tested device as faulty) or press
START
to repeat this test (e.g. when im-
proper result was caused by an error in connections).
- For R
ISO
measurement, there are three options:
- no R
ISO
measurement,
R
ISO
1
- R
ISO
standard
measurement,
R
ISO
3
- R
ISO
triple measurement - only for welding machines.