82
6.7.5 Counting function
The reference weight can be determined
using the following function keys:
>
Function key F1 Ref 10
Reference weight determination from a preassigned reference sample quantity. Place 10 counted
pieces on the scale, then press the F1 Ref 10 key. The scale determines the reference weight and
displays "10 pieces. Parts that have been placed further on the scale are counted continuously. The
standard reference sample quantity "10" can be selected in the Setting mode / Program settings /
Counting under Standard reference pieces. (Chapter 6.7.14).
If, for example, 25 is stored there, the new function key is F1 Ref 25 and 25 pieces must be loaded to
determine the reference weight.
>
Function key F2 Ref Variab
Reference weight determination from a freely selectable reference sample quantity. Place a certain
number of reference parts on the scale. Press the F2 Ref Variab key and enter the quantity (1 to 9999)
of the parts on the scale using the numeric keypad. Press the F6 Enter key. The terminal determines
the reference weight and displays the number of pieces on the scale. Parts that have been placed
further on are counted consecutively.
>
Function key F3 Ref Weight
Enter a known reference weight. Press the F3 Ref weight key and enter the known reference weight
using the numeric keypad. An existing reference weight is displayed and can be overwritten. By
default, the reference weight is entered in grams. If you want to enter the reference weight in
kilograms, you can change the unit by pressing the key . Confirm your entry with F6 Apply.
The terminal has accepted the reference weight and displays the quantity loaded.
New reference weight
Existing reference weights are generally overwritten by a new reference weight determination or by a fixed
value being transferred from the fixed memory.
Displaying the reference weight
Press the F3 Ref weight key or activate the auxiliary display (see chapter 5.10).
End counting
The counting process is terminated by deleting the reference weight. The reference weight can be deleted by
pressing the F3 Ref Weight and F2 Delete keys or simply by pressing the key . The display shows the
current weight value.
Read-only memory Reference weight (Ref Fix)
You can store 999 counted items in a database that is protected against power failure.
Each memory location has the following entries:
>
Three-digit sequential number
>
Designation (alphanumeric max. 18 characters)
>
Article number (alphanumeric max. 18 digits)
>
Reference weight
>
Nominal piece count
>
Tolerance minus
>
Tolerance plus