background image

 

 

 

103 

Editing an existing fixed value from the list 

Edit the selected fixed value via F6 Edit. 
Parameter Name is highlighted, call via F6 Edit, input via keyboard.  
Accept the entry with F6 (abort without accepting the entry with F5 Abort). 
Further parameters are edited in the same way. Only the same number of decimal places can be used for the 
tolerance fields Minus and Plus. 
 

Deleting an existing fixed value 

Scroll with the arrow keys F3 and F4 or call up via the article number or the sequential number. 
Delete the called fixed value with F2 Delete. Answer the confirmation prompt with YES. 
The resulting gap in the consecutive numbering is closed. 
 

Save new fixed value 

Press F1 New. 
The memory location number is assigned by the terminal. The first number in brackets indicates the number 
of free memory locations, the second number the occupied memory locations. 
Parameter Name is highlighted, call via F6 Edit, input via keyboard.  
Accept the entry with F6 (abort without accepting the entry with F5 Abort). 
Further parameters are edited in the same way. Only the same number of decimal places can be used for the 
tolerance fields Minus and Plus. 
 

Exiting the list Fixed values 

Press the F5 Back key to exit the list. 
 
 

b) Calling up the list of fixed values of control weights for checking: 

In Level 1 Control Functions, press the F2 Contr. fix key. 
The display shows the last fixed value called up. 
 

Selection within the list 

 

Scroll with the arrow keys F3 or F4 until the desired memory location is reached,  
Accept the desired fixed value with F6 Accept 

 

Direct call via article number or memory location number 
Use the numeric keypad to enter the item or memory location number. 
(Determining whether the article or memory location number is called up in setting mode, entering the 
article number as a search criterion is left-justified, entering the memory location number as a search 
criterion is right-justified) Accept the desired fixed value with F6 Accept 
 

When the fixed value is called up, setpoints and tolerances for checking are also accepted, if saved with .  
Press the CL key to delete the read-only memory from the working memory. 
To delete all stored reference weights, use the "Delete reference weights" function in the setting mode.  
 

Note 

If the function read-only memory call via barcode is activated and one is called that is not contained in the 
fixed memory, the error message "Error 83 - Fixed value not available" appears. You return to the weighing 
mode and the selected function is deleted.  
 

Summary of Contents for 3025

Page 1: ...www soehnle professional com Operation Manual Terminal 3025 ...

Page 2: ... use could result in damage to you or the terminal If you have any questions or have problems with your terminal that are not covered in this manual please contact your Soehnle Industrial Solutions service center or Soehnle Industrial Solutions Customer Service Telephone 49 7191 3453 220 Fax 49 7191 3453 211 E Mail info sis gmbh For more information and documentation see https www soehnle professi...

Page 3: ...ing up the setting mode 17 4 2 Program settings 18 4 3 Setting mode Terminal settings 19 4 4 Setting mode Scale settings 25 4 5 Setting mode Communication settings 28 4 6 Setting mode Service settings 37 4 7 Reset terminal 37 5 BASIC FUNCTIONS 38 5 1 Switch on 38 5 2 Switch off 38 5 3 Program selection 38 5 4 Set to zero 38 5 5 Tare function 39 5 6 Gross display 39 5 7 Scale switching 39 5 8 Tenfo...

Page 4: ...6 Selection options for the totals display 73 6 6 7 Relief factor for totalising 73 6 6 8 Auto total pick 73 6 6 9 Reference scale for totalisation 73 6 6 10 Assigning and deleting the sequential number 74 6 6 11 Item counter 74 6 6 12 Print triggering keys 74 6 6 13 Display of current totals 74 6 6 14 Display individual items 75 6 6 15 Display options with the Info key 76 6 6 16 Transfer of curre...

Page 5: ...are function 101 6 8 5 Control function 101 6 8 6 Totalising and Picking 104 6 8 7 Display options with the Info key 108 6 8 8 Transfer of current values to the fixed memory 108 6 8 9 Organizational data Identifier 109 6 8 10 Settings in the setting mode 109 7 ERROR MESSAGES 113 8 ADDITIONAL INFORMATION 113 ...

Page 6: ...erminal The manufacturer is not liable if the following instructions are not observed If electrical components are used under increased safety requirements the relevant regulations must be observed Never service the terminal while it is live The warranty does not apply in the event of improper installation The Terminal 3025 must not be used in hazardous or explosion endangered areas The permissibl...

Page 7: ...ent Replaced parts shall become the property of Soehnle Industrial Solutions If the rectification of the defect or replacement delivery fails the statutory provisions shall apply The warranty period begins on the day of purchase Please keep the invoice as proof If service is required please contact your dealer or Soehnle Industrial Solutions customer service In particular no warranty is given for ...

Page 8: ...instrument whose performance can be affected by unfavourable environmental factors After unpacking or transport acclimatize for 2 hours at room temperature The permissible ambient temperature is 10 to 40 C The place of installation is a decisive factor for the functioning of your scale Protect the terminal as far as possible from shocks vibration strong heat or cold temperature fluctuations draugh...

Page 9: ...Interfaces 1 x RS 232 or USB in the basic version Internal with R232 USB and Ethernet 2 additional interface slots RS232 Option IO card for 6 outputs and 4 inputs freely assignable Working temperature 10 C to 40 C 1 analogue measuring point in the basic version 1 additional analogue measuring point can be connected internally Calibratable resolution 10 000 000 e internal resolution 16 000 000 d Ve...

Page 10: ...ctive Symbols in the information field of the display Below tolerance Within tolerance Setpoint reached exactly Above tolerance Under minimum load Above upper limit class 5 Symbol for standstill After the sample has been placed on the scale it needs time to settle As long as no stability has been reached the dimension symbol kg g is hidden If stability has been reached and a stable value is availa...

Page 11: ...e center or your local calibration office Do not under any circumstances damage the official seals otherwise the calibration validity will expire 2 4 1 Nameplate The terminal has an electronic nameplate Display after pressing the Info button Brutto Netto F 1 F 2 F 3 F 4 F 5 F 6 01 Gewichtswert Dimensions zeichen Max Min e SNR XXXX XX XXXX Eichz XXXX IDENT Chip 2 4 2 Calibration counter The calibra...

Page 12: ...ght value is then displayed with the previously stored zero point If you remove the cause of the underload or overload the scale will go to zero and is ready for operation without having to be switched on again Backlight and colour Can be adjusted in Setting mode Terminal Display The default setting is 80 for the backlight and white for the background colour Behaviour in case of power failure The ...

Page 13: ... Weighing taring Version Scale parameter Alibi memory for Totalising and picking Display Allocation table measuring points Interface 1 RS232 internal service Counting Keypad Reference points for calibration Interface Port 2 Specialists Checking Date Time Internal measuring point Interface Port 3 only I O port programming Organizational data USB Error memory Ethernet User password Print template Lo...

Page 14: ...ls takes place rolling with the function change key Navigation with the function keys Move the marker up in the list Move the marker down in the list End Exiting the setting mode Continue Call of the marked menu item Back Return to superordinate level Cancel Return without accepting a change Delete Deleting an item with a return jump Accept Acceptance of an entry or change Cursor to the right Curs...

Page 15: ...individually under Terminal Display Can be switched on and off by pressing the F3 key Symbol bar Shows information on the status of the connected measurement loop Information field Shows organizational data totals during totalising special control characters classes during classification operator guidance switching points depending on situation or individually adjustable Display field Displays wei...

Page 16: ...e structure Setting mode Program settings Terminal Scale Communication Service Weighing taring Version Scale parameter Alibi memory for Totalising and picking Display Assignment table for measuring points Interface 1 RS232 internal service Counting Keypad Reference points for calibration Interface Port 2 Specialists Checking Date Time internal measuring point Interface Port 3 only I O port program...

Page 17: ...the selection menu Setting mode Navigation in Setting Mode The function keys F1 to F6 are available for controlling the functions They assume different functions depending on the program and situation The respective function is shown on the display in the function bar Meaning of the function keys Move the marker up in the list Move the marker down in the list End Exiting the setting mode Continue ...

Page 18: ...4 2 Program settings In the Program Settings setting mode you can adapt the user programs of your scale to your specific requirements How to activate the application programs and their typical applications are described in Part 6 Application Programs 1 Weighing taring Settings and parameters see chapter 6 5 16 2 Totalising and picking Settings and parameters see chapter 6 6 18 3 Counting Settings ...

Page 19: ...ion Latin 1 Print parameter Output of all setting parameters of the terminal incl the currently connected measuring point via the interface with the use EDP 2 unidirectional 2 Display Backlight Input Setting luminosity input in percent via numeric keypad 30 is darker 100 is full luminosity Factory setting 80 Background colour Selection Background colour setting green white light blue yellow blue m...

Page 20: ...e Factory setting None 3 Keypad Keypad key Single multiple or all keys can be locked on the keyboard By pressing the key to be locked the status of the key can be changed from open to locked and vice versa Press the Enter key to save and exit for 3 seconds Key tone Selection off or on Keyboard confirmation beep Factory setting off Validation tone Selection off or on Receipt beep e g error message ...

Page 21: ...Description ORG A Input Name of the organizational data memory A Numbers or plain text via keyboard e g Article no max 12 characters Description ORG B Input Name of the organizational data memory B digits or plain text via keyboard max 12 characters Description ORG C Input Name of the organizational data memory C digits or plain text via keyboard max 12 characters Description ORG D Input Name of t...

Page 22: ...reate new fixed value Press F1 New The memory location number is assigned by the terminal The first number in brackets indicates the free memory locations the second number the occupied memory locations Parameter number is marked call via F6 Edit input via number alpha keyboard Accept the entry with F6 Cancel without accepting the input with F5 Cancel Parameter entry is marked call via F6 Edit ent...

Page 23: ...d gives access again 8 Login function Activation Selection on or off Factory setting off Password request Selection on or off Factory setting off Create change password Input of the internal terminal password System alphanumeric via the numeric keypad Password with up to 35 characters After activation access is via a password query e g after switching on the terminal or after logging off by pressi...

Page 24: ...prompt Import comes with YES or NO If NO is selected no data is loaded Export The current user settings are saved on the micro SD card by accepting F6 Continue Export for autom import The current user settings are saved on the micro SD card and the setting parameters are automatically reloaded in the event of a software problem ...

Page 25: ...le in the service area Livestock scale Calm Normal Troubled Factory setting Service setting The selection function is only possible with the release user filter setting in the service area Hold mode Selection not active no hold function Still key Hold at stability outside zeroing range Cancel by pressing the power button Still empty Hold at standstill outside zeroing range Cancel by relieving the ...

Page 26: ...ual always refers to the Total tare setting Factory setting off Second unit Selection g t lb mg oz Can be activated as auxiliary display function in the display via key F3 Auxiliary display in weighing mode In the selection the adjusted unit is not displayed as the first unit Empty message Input of the percentage of the weighing range in which the empty message is displayed Enter the percentage us...

Page 27: ... measured value Query of the adjusted raw measured value from the zero point Current raw measured value Query of the current raw measured value variable depending on the load on the scale Reference point 1 weight value Query of the weight value from the 1st support point Reference point 1 raw measured value Query of the adjusted raw measured value from the 1st support point Reference point 2 weigh...

Page 28: ...ingle input Search Single entry via input consecutive number Find function Search Multiple entries by entering oldest and youngest consecutive number Search additional field Search Single entry via input contents additional field Assignment of additional data Selection None Barcode Org data A Org data B Org data C Org data D Date Time Price labelling special software 2 Interface 1 RS232 internal F...

Page 29: ...Do not execute Individual settings on the interface are retained Execute Resets interface to factory setting 3 Interface Port 2 Further details see also interface description 470 508 090 Usage of serial interface Selection none 2795 14 2795 12 2795 20 Data processing EDP 2 unidirectional Barcode Second large display Digital scale If the device is already used via another interface output this sele...

Page 30: ...ce to factory setting 4 Interface Port 3 Further details see also interface description 470 508 090 Usage of serial interface Selection none 2795 14 2795 12 2795 20 Data processing EDP 2 unidirectional Barcode Second large display Digital scale If the device is already used via another interface output this selection position is displayed with disabled Factory setting none Baud rate Selection 1200...

Page 31: ... logger If the device is already used via another interface output this selection position is displayed with disabled Factory setting none 6 Ethernet Further details see also interface description 470 508 090 Use of Ethernet Selection none Data processing If the device is already used via another interface output this selection position is displayed with disabled Factory setting none IP Address In...

Page 32: ...s empty Factory setting on Decimal separator Selection Comma dot or no Factory setting Decimal point Thousands separator Selection Comma dot or no Factory setting none Hide unit Selection on or off Off Unit is not suppressed Factory setting off Consecutive number print key Input The number for the first print image can be entered here If the value zero is entered the function is not active As of n...

Page 33: ...n off or on Factory setting off with one the data record with the print image I is sent automatically via the EDP interface after each standstill design of print image I via service software required EDP mode unidirectional Selection for EDP mode unidirectional terminal automatically triggers the selected data set none A Send data record 1 x immediately B Send data record 1 x at standstill weight ...

Page 34: ...ode Control character numeric via numeric keypad Factory setting 0 First position filter no 01 Input Start filter position numeric via numeric keypad Factory setting 1 Last position filter no 01 Input End of filter position numeric via numeric keypad Factory setting 70 Multiscan number Input of the number of multiple scans from 0 to 4 using the numeric keypad Factory setting 0 Code 1 Scan ORGA Inp...

Page 35: ...filter for scan ORGC Filter for scan ORGD Like filter for scan ORGA valid for filter for scan ORGD 10 Batch processing further details see also manual 3025 Chapter 7 10 Batch code 1 Input Control commands from interface description 3025 document 470 508 090D Factory setting 0 Batch code 2 Input Control commands from interface description 3025 document 470 508 090D Factory setting 0 Batch code 3 In...

Page 36: ...e pause between the batch codes numeric via numeric keypad Factory setting 100 ms Pause between batch code 4 5 Input Time for the pause between the batch codes numeric via numeric keypad Factory setting 100 ms Pause after end Input Time for the pause at the end numeric via numeric keypad Factory setting 100 ms 11 IO control further details see also manual 3025 Chapter 7 11 IO mode Selection off on...

Page 37: ...ly be maintained by trained service personnel 4 7 Reset terminal To reset all settings except scale parameters and service area of the terminal back to the factory settings you can perform a reset of the terminal with the selection in the Setting mode Terminal Reset Terminal and press the function key F6 Continue ...

Page 38: ... arrow key F4 down arrow key Start the selected program Press F6 to confirm For information on all user programs see Chapter 6 5 4 Set to zero Zeroing corrects small deviations from the zero point e g due to contamination of the scale Press the zeroing key The following character appears in the toolbar when zero is displayed Zero range calibratable and non calibratable 1 to 3 of weighing range If ...

Page 39: ...S232 interface with measuring point no 30 Switching between the connected measuring points can be done manually or automatically Manual switching Press the button briefly The connected measuring points are connected one after the other Press and hold the button long An input field opens the measuring point number can be directly entered and accepted Automatic switchover For description see program...

Page 40: ...ds Non approvable scale If the scale is not calibrated for legal metrology the resolution is ten times higher and appears constantly in the display Pressing a key switches on the tenfold higher resolution Pressing the key again switches the tenfold higher display off again As long as the tenfold higher display is active X10 appears in the toolbar The function is deactivated when the device is swit...

Page 41: ...new hold cancel by switch on button Max empty Hold at maximum reached value with stability outside empty signal higher weight sets new Hold lift by relieving the scale Drag key Hold at each increase of the weight without stability outside empty message higher weight sets new hold Cancel by pressing the power button Drag Empty Hold at every increase of the weight without stability outside empty sig...

Page 42: ...display Setting the auxiliary display To set the function of the auxiliary display first switch to the setting functions level with the function change key Gross x 10 Auxiliary display Prog ram Setting mode F 1 F 2 F 3 F 4 F 5 F 6 Press the F6 Setting Mode key Use the arrow keys F3 and F4 to move to the Terminal item in the list and select it by pressing the F6 Continue key Now move to Display wit...

Page 43: ...e auxiliary display The following function states can be set in the submenu auxiliary display Second unit Tare value x10 Resolution Gross net Current weight Reference quantity Reference weight Setpoints Difference to setpoint Signal light Note the setting of the function must correspond to the user program The factory setting of the auxiliary display is the second unit ...

Page 44: ...37 g Milligram mg 0 001 g Ounce oz 28 349523125 g To select the second unit use the Function change key to switch to the setting functions level Gross x 10 Auxiliary display Prog ram Setting mode F 1 F 2 F 3 F 4 F 5 F 6 Press the F6 Setting Mode key Use the arrow keys F3 and F4 to move to Scale in the list and select Continue by pressing the F6 key Then press the F6 Continue key to select Scale pa...

Page 45: ...45 You can move in the list with the arrow keys F3 and F4 Press F6 Apply to define the second unit After selecting the second unit press F1 End to return to the weighing mode ...

Page 46: ...available for calling up the organizational data memory Input Designation First enter the names max 12 characters for the required organizational data memories in the setting mode Call up the setting mode by pressing the F6 key in the Setting functions key level Use the arrow key to highlight F4 Terminal F6 Next Use the arrow key F4 to select Organisation data F6 Continue Designation A ORG to D OR...

Page 47: ...s the name for the organ data memory A In Fixed Values Org Data A enter the articles to be entered such as Screws Nuts Washers Dowels etc as an entry As designation for the org data memory B operator has already been created Now enter the possible operators such as Schmidt Müller Meier Krause etc as entries in Fixed values Org data B The fixed values can now be activated from the database at any t...

Page 48: ...rd Accept the entry with F6 Cancel without accepting the entry with F5 Cancel Exiting the database Press F5 back to exit the list Database maintenance via optional Soehnle Professional user software 3025 For the requirement of a simple and multiple application to the care of the Org data base for the designations and contents there is the PC user software 3025 The user software is connected to the...

Page 49: ...ecimal point setting mode The input field can only be filled up to the specified number of characters further entries are not displayed and are not accepted Alphanumeric fields The digits 0 to 9 can be entered directly using the numeric keypad The entry is left justified The numeric keys are also assigned uppercase letters and special characters in multiple assignments The special character keys a...

Page 50: ...Cancel key lock Call up Terminal Keypad Keypad keylock in the setting mode The keyboard is shown schematically in the display Locked keys are marked with a cross in the display By pressing the key the status of the key can be changed from locked to open and vice versa Accept and activate the change by pressing the Enter key for 3 seconds If no more keys are locked the symbol for key lock disappear...

Page 51: ... Enter key for 3 seconds Logout or end by pressing the C button for 3 seconds Details for the dialog and the commands of the EDP data communication can be found in the interface description 470 508 090 Activating and Setting In the Terminal Login function setting mode different variants can be user defined ranging from a simple process start query to a password query with monitoring by an external...

Page 52: ...ord with the logon rejection is sent then the terminal is in weighing mode Activation of password query external EDP query Call up setting mode by pressing the F6 key in the setting functions key level Use the arrow key to highlight F4 Terminal F6 Next Use the arrow key to highlight F4 Login function F6 Next Use the arrow key F4 to highlight the password request F6 to proceed to the password reque...

Page 53: ...th the command code is sent to the computer for verification and release and the hourglass appears in the display Within 5 seconds the computer must release the password if the password release does not come within this time the security question Login Cancel appears Or if an incorrect password is confirmed display message Error 02 Incorrect password the password query is repeated If the F5 Cancel...

Page 54: ...ate change password Enter the desired password with max 35 digits and confirm with F6 Comments After entering an internal password there is no longer an external EDP password query When creating an internal terminal password the password is always requested in the Login function area If the password is forgotten the Login function area can be accessed by entering CLEAR without a password After the...

Page 55: ... F6 Next Use the arrow key to highlight F4 Login function F6 Next Password prompt internal terminal Enter the password and confirm with the F6 key If the password is forgotten you can access the Login function area by entering CLEAR without a password Use the arrow key F4 to highlight Delete password F6 Continue The internal password is deleted Enter login text The text for the Login query window ...

Page 56: ...D card data backup menu the current user settings from the micro SD card are reloaded into the terminal by accepting F6 Continue The security question Import with the function keys F1 to F3 YES the data transfer from the micro SD card to the terminal is executed with the function keys F4 to F6 NO no data transfer from the micro SD card to the terminal takes place Export In the SD card data backup ...

Page 57: ...icking Counting Checking I O port programming 6 1 Selection of application programs Switch on the terminal After the power on routine the terminal will automatically start the application program that was last activated To select another application program use the function change key to switch to the Setting functions level 6 2 Function keys Level Setting functions Gross x 10 Auxiliary display Pr...

Page 58: ...t Move the marker down in the list End Exiting the setting mode Continue Call of the marked menu item Back Return to superordinate level Cancel Return without accepting a change Delete Deleting an item with a return jump Accept Acceptance of an entry or change Cursor to the right Cursor to the left 6 4 Special programs In addition to the standard user programs it is also possible to program sequen...

Page 59: ...re Selection of options in setting mode scale menu 6 5 1 Function keys The function keys F1 to F6 are available for controlling the functions These are assigned in 2 levels the change between the levels is made by the function change key 1 Level Tare functions Tare fix Add tare Prov tare Multi tare Accept fixed F 1 F 2 F 3 F 4 F 5 F 6 2 Level Setting functions Gross x10 Auxiliary display Prog ram ...

Page 60: ... you can set which data are displayed in the information field The following options are available Factory default Change between the individual functions used e g switching functions outputs and inputs and organizational data memory IO control Continuous display of output and input states Organizational data database Continuous display of the organizational data memory Fixed tare values Continuou...

Page 61: ...out making any changes If you want to change the value of the tare memory you can enter a new value using the numeric keypad and store it in the tare memory as a manual tare entry by pressing F6 6 5 6 Taring with manual tare input Taring a known container weight e g when the container on the scale is full Enter the known tare weight using the numeric keypad and confirm your entry by pressing the F...

Page 62: ...emory you can enter a new value using the numeric keypad and store it in the tare memory as a manual tare entry by pressing F6 6 5 8 Additive tare Taring of several even different containers with known tare weight Press the F2 Add Tare key in the Tare functions level Enter the known tare weight of the first container using the numeric keypad Confirm with the F6 Enter key Repeat the procedure for t...

Page 63: ... first weight value is stored in the tare memory and subtracted from the weight display The weight is displayed with the message Net During automatic taring the message Auto tare active appears If the scale is unloaded the tare memory is automatically cleared and the message Auto tare active appears in the display The scale returns to the gross display You can delete auto tare with the scale loade...

Page 64: ...lue Parameter Name is highlighted call via F6 Edit input via numeric alpha keyboard Input with F6 Accept abort without accepting the input with F5 Abort Parameter Tare value is highlighted call via F6 Edit input via digits alpha keypad Input with F6 Accept abort without accepting the input with F5 Abort Delete an existing fixed value Scroll with the arrow keys F3 and F4 Delete the marked fixed val...

Page 65: ...nd one is called that is not contained in the fixed memory the error message Error 83 Fixed value not available appears You return to the weighing mode and the selected function is deleted Delete Fixed tare value To delete the tare value and the designation from the working memory press the CL key 6 5 12 Error messages during taring Error 06 Taring not possible Possible causes Empty scale Scale in...

Page 66: ...firm Fix the value can be stored directly in the Fix memory The input window for the fixed tare values appears To enter the designation of the fixed tare value press the F6 Edit key Enter the name using the keypad and confirm your entry by pressing the F6 Enter key 6 5 15 Organizational data Identifier Detailed information on the organizational data can be found in chapter 5 11 on page 46 ...

Page 67: ...row keys F3 and F4 Delete the marked fixed value with F2 Delete The resulting gap in the consecutive numbering is closed Save new fixed value Press F1 New The memory location number is assigned by the device The first number in brackets indicates the number of free memory locations the second number the memory locations occupied by fixed values Parameter Name is highlighted call via F6 Edit Input ...

Page 68: ...nd inputs and organizational data memory IO control Continuous display of output and input states ORG data database Continuous display of the organizational data memory Fixed tare values Continuous display of the activated fixed tare value memory Show barcode Continuous display of the scanned barcode Code128 Weight value Display bar code128 with the current stability weight value Factory setting F...

Page 69: ...the following the term totalising is used as an example Totalising memory for several connected scales A central totalising memory is maintained for all connected weighing instruments Totalising memory capacity A maximum of 999 bookings are possible in the total memory The totalising memory must then be erased 6 6 1 Function keys The function keys F1 to F6 are available for controlling the functio...

Page 70: ... set to factory setting Hint To additionally display the 10 times higher resolution x10 or the tare weight for example you can activate an additional auxiliary display by pressing the F3 key You can set the value to be displayed in the auxiliary display in the Setting mode under Setting mode Terminal Display Auxiliary display Illustration auxiliary display with x10 display ...

Page 71: ...ay of the totalising memory Last item Continuous display of the last item IO control Continuous display of output and input states ORG data database Continuous display of the organizational data memory Fixed tare values Continuous display of the activated fixed tare value memory Show barcode content Continuous display of the scanned barcode Code128 Weight value Display bar code128 with the current...

Page 72: ...Weight values are added to the weight total piece numbers are added separately to the piece total Cancellation F4 You can use the F4 Cancel key to reverse or reweigh individual items After pressing the Cancellation key you are prompted either to debit the scale with the weight to be cancelled or to enter the serial number of the weight to be cancelled In advance you can enter information on the cu...

Page 73: ...alising memory and is displayed with the error message Error 55 Place weight on scale In this way you can prevent for example duplicate entries being made for one item 6 6 8 Auto total pick The auto total function can be activated in the setting mode Then only the first position must be stored with the Plus key all other positions are automatically stored in the totalising memory after they have b...

Page 74: ... cancellation the transaction counter is reduced by 1 The transaction counter is always automatically set to zero when the total is deleted 6 6 12 Print triggering keys Additional keys or functions such as plus manual cancellation and total can be used in the totalising picking function as print triggering keys e g for document printing The prerequisite is that a print image has been stored for th...

Page 75: ...memory by pressing and the F5 Plus key one after the other The item list starts with the most recent entry Use the arrow keys F3 older entries and F4 newer entries to cycle through the list The last 50 entries can be called up Press the F5 Back key to exit the display of the individual items and return to the previous display Note Display information field with item list In the Setting Mode Progra...

Page 76: ...talising memory F5 Plus Displays all items stored in the totalising memory item list The individual items can be called up using the arrow keys F3 F4 Tare key Displays the current tare weight in the tare memory After pressing the key press the key 6 6 16 Transfer of current values to the fixed memory To store fixed tare values directly in the fixed memory first store the tare value or enter the ta...

Page 77: ...g mode key in the Setting functions level and then press F6 Apply to call up the program settings F6 Next calls up the program settings for the selected application program 2 Totalising and Picking Totalising or Picking Selection of totalling or picking for the following settings Totalising means that the individual items are removed from the scale after they have been stored in the totalising mem...

Page 78: ...ration can be started Factory setting Display Print Clear Clip off factor Input in d via numeric keypad The number of digits by which the scale scale must be unloaded during totalisation so that you can store weights in the totalising memory again If the weight is not unloaded it will not be stored in the totalising memory error message Error 55 Place weight on scale You can enter values between 0...

Page 79: ...s F1 to F6 are available for controlling the functions These are assigned in 4 levels the change through the levels takes place via the function change key 1 Level Counting functions Ref 10 Ref Variab Ref Weight Total Clear Plus F 1 F 2 F 3 F 4 F 5 F 6 2 Level Control functions Ref Stat Ref fix F 1 F 2 F 3 F 4 F 5 F 6 3 Level Tare functions Tare fixed Add tare Prov tare Multi tare Accept fixed F 1...

Page 80: ...e x10 resolution or the tare weight you can activate an additional auxiliary display by pressing the F3 key You can set the value to be displayed in the auxiliary display in the setting mode under Setting mode Terminal Display Auxiliary display Illustration auxiliary display with reference quantity ...

Page 81: ...he last item ORG data database Continuous display of the organizational data memory Fixed values for reference weight Continuous display of the reference weight memory activated Show barcode Continuous display of the scanned bar code Code128 Weight value Display bar code128 with the current stability weight value or counter value Example Setting the display of the information field to fixed values...

Page 82: ...n key F3 Ref Weight Enter a known reference weight Press the F3 Ref weight key and enter the known reference weight using the numeric keypad An existing reference weight is displayed and can be overwritten By default the reference weight is entered in grams If you want to enter the reference weight in kilograms you can change the unit by pressing the key Confirm your entry with F6 Apply The termin...

Page 83: ...e called fixed value with F2 Delete Answer the confirmation prompt with YES The resulting gap in the consecutive numbering is closed Save new fixed value Press F1 New The memory location number is assigned by the device The first number in brackets indicates the number of free memory locations the second number the occupied memory locations Parameter Name is highlighted call via F6 Edit input via ...

Page 84: ... the Counting functions level Totalising means that the individual positions are weighed one after the other the weight in the totalising memory is added and the scale is unloaded again after each individual weighing Picking means that the weighed positions remain on the scale The weight is added to the total memory and the scale returns to zero without being unloaded It is not permitted to change...

Page 85: ...eigh individual items After pressing the Cancel key you are prompted either to debit the scale with the weight to be cancelled or to enter the current number of the weight to be cancelled Information on the current number via the key combination Info and Plus F5 Enter the current number of the item to be cancelled using the numeric keypad and confirm with F6 Accept If you do not enter a sequential...

Page 86: ...y and the error message Error 55 Place weight on scale appears In this way you can prevent for example double storage of an item Auto Totalise The Auto Totalise function can be activated in the setting mode In this case you only need to confirm the first position by pressing the Plus key All other positions are automatically added after weight has been reached and stability is stored in the totali...

Page 87: ... triggering keys The Plus Reversal and Total keys can trigger document printing The prerequisite is that a print screen has been defined for the respective key The Display print and delete option must be selected for the total key in the Total key function setting mode Display current totals Within a total you can display the total values achieved by pressing the key and then the F4 Sum key one af...

Page 88: ...the target quantity and the permitted tolerance in the piece check mode Further information can be found under Check piece Totalising of different counted parts You can prevent the totalising of different parts by selecting Clear total in the Setup mode to allow a new reference weight to be determined only after the totalising memory has been cleared 6 7 7 Reference statistics Determination of the...

Page 89: ...ypad to enter the desired piece count confirm with the F6 Apply Enter the accepted tolerance in pieces using the numeric keypad and confirm with the F6 Enter key If you want to define the plus and minus tolerance separately activate in the setting mode under Program settings Counting Tolerance separated The display switches to the Counting display view the traffic light function with the special c...

Page 90: ...e frozen status released Activate the start mode in the setting mode under Program settings Counting Start mode End piece control Call up the 2nd level control functions via the key and press key F5 Quantities contrast then delete via key F2 Delete or complete deletion via key 6 7 9 Counting with several weighing bridges If several scales are connected the reference weight can be determined and qu...

Page 91: ...to 5 d the scale will accept 5 g as the minimum weight for the total weight of the reference parts If you increase the factory setting to 20 d the scale will assume a minimum weight of 20 g for the total weight of the reference parts Optimization Ref reference weight optimization The reference weight of a batch determined by the reference weight calculation from e g 10 or 20 parts is usually not t...

Page 92: ...f the scale is to be optimized Example 0 off 50 30kg for a scale with 60kg maximum load 100 60kg for a scale with 60kg maximum load Example of reference weight updating ADD Mode If the minimum reference weight is not reached the parts are requested until the minimum reference weight is reached The scale calculates the reference weight after the requested parts add xx pieces have been placed on the...

Page 93: ...list The individual items can be called up using the arrow keys F3 F4 Tare key Displays the current tare weight in the tare memory After pressing the key press the key 6 7 12 Transfer of current values to the fixed memory To store reference weights and target values directly in the fixed memory for piece checking first determine the reference weight and add the target value piece with tolerances P...

Page 94: ...6 Edit key Enter the name using the keypad and confirm your entry by pressing the F6 Apply key Enter the article number in the same way 6 7 13 Organizational data Identifier Detailed information on the organizational data can be found in chapter 5 11 on page 46 ...

Page 95: ...When the ADD mode is activated the scale requests the missing parts until the minimum reference weight is reached If the calculated reference weight is less than 10 of the smallest digit of the scale the error message Error 04 Reference weight too small appears Reference weight optimization Enter percentage The scale optimizes the reference weight when additional parts are loaded The upper limit o...

Page 96: ...n means that the totalising memory must first be cleared before a new reference weight can be determined This prevents different parts from being added together Factory setting off Total or picking Select Totalise or picking for the following settings Totalising means that the individual items are removed from the scale after they have been stored in the totalising memory Picking means that the in...

Page 97: ...Reference weight Selection of existing fixed values from the list Scroll with the arrow keys F3 and F4 Call the desired fixed value with F6 Edit Edit an existing fixed value from the list Press F6 to call up editing of the selected fixed value Parameter Name is marked call via F6 Edit Input via keyboard Accept the input with F6 Cancel without accepting the input with F5 Cancel Further parameters a...

Page 98: ... Display information field Selection of which data is to be displayed in the information field Counting Toggle between the individual functions used counting with the reference weight and weight value in gross net e g control light totalising memory Control display Continuous display of the traffic light function Total Continuous display of the totalising memory Last item Continuous display of the...

Page 99: ... shown in the display by a control light with the colours yellow green red including the respective range symbol which visualizes the deviation from the pre set setpoint value Control ranges under tolerance Within tolerance exactly reached with setpoints Above tolerance 6 8 1 Function keys The function keys F1 to F6 are available for controlling the functions These are assigned to three levels the...

Page 100: ...erence to the setpoint you can activate an additional auxiliary display by pressing the F3 key The value to be displayed in the auxiliary display is set in setting mode under Setting mode Terminal Display Auxiliary display Illustration auxiliary display with difference to setpoint value ...

Page 101: ...n tolerance 3 Above tolerance Target values reached exactly 6 8 4 Tare function For a description and detailed instructions of the tare function refer to chapter 6 5 on page 59 6 8 5 Control function Input of setpoint values In Level 1 Control Functions press the F1 Setpoint key You can now place a sample part on the display to store the setpoint or enter the setpoint in the adjusted unit using th...

Page 102: ...irectly by pressing the key Check fixed values contr fix You can store up to 999 setpoints in a power failure proof database Each memory location has the following entries Three digit sequential number Designation alphanumeric max 18 characters Article number alphanumeric max 18 digits Setpoint Tolerance minus Tolerance plus a Call the list of fixed values for control weights to edit the fixed val...

Page 103: ...iting the list Fixed values Press the F5 Back key to exit the list b Calling up the list of fixed values of control weights for checking In Level 1 Control Functions press the F2 Contr fix key The display shows the last fixed value called up Selection within the list Scroll with the arrow keys F3 or F4 until the desired memory location is reached Accept the desired fixed value with F6 Accept Direc...

Page 104: ...e tolerance 6 8 6 Totalising and Picking This function is also available in control mode With this function control goods can be totalled picked The totalising functions are controlled with function keys on the Control functions level Totalising means that the individual positions are weighed one after the other the weight in the totalising memory is added and the scale is unloaded again after eac...

Page 105: ...he error message Error 46 Cancellation or back weighing are indicated in the print document by the name Cancellation The consecutive number is incremented by one during cancellation the transaction counter is reduced by one If the transaction counter is set to 0 no further cancellation is possible indicated by error message Error 48 Total F4 or with the CL key By pressing the F4 Total key the tota...

Page 106: ... assigned a four digit sequential number The sequential number is not reset in the factory setting Optionally the consecutive number can be reset in the setting mode not active no reset factory setting deleting the totalising memory Switching off the terminal Function key Delete F1 current no when calling via keys and F4 Total Item counter The item counter only counts the actually totalled transac...

Page 107: ...y by pressing the key and the key F5 Plus one after the other The item list starts with the most recent entry Use the arrow keys F3 older entries and F4 newer entries to cycle through the list The last 50 entries can be called up Press the F5 Back key to exit the display of the individual items and return to the previous display Note Display information field with item list In the Setting mode Pro...

Page 108: ... Display of stored fixed values Controlling Contr Fix Memory F3 Setpoint Display of the active setpoint with tolerances F4 Total Display of the current totals in the totalising memory F5 Plus Displays all items stored in the totalising memory item list The individual items can be called up using the arrow keys F3 F4 Tare key Displays the current tare weight in the tare memory After pressing the ke...

Page 109: ...ss the F6 Setting mode key in the Setting functions level and call up the program settings there with F6 Apply F6 Next calls up the program settings for the selected application program 4 Checking tolerance separated Selection off or on Selection on causes that and tolerance for the piece control can be defined and queried separately Selection off causes the tolerance to be queried only once and i...

Page 110: ...is to be reset Not active Clear total memory On Off Function key activates a separate function key Factory setting not active Total key function Selection of which functions are to be linked to the sum key Display totals only After pressing the function key Total the total is displayed The user can then decide whether the total memory is to be deleted or not Press the F2 key to clear the total mem...

Page 111: ...second number the memory locations occupied by fixed values Parameter Name is highlighted call via F6 Edit input via keyboard Accept the entry with F6 Accept Cancel without accepting the input with F5 Cancel Further parameters are processed in the same way Exiting the list Fixed values Press the F5 Back key to exit the list Call fixed value Selection of whether the fixed value is called up via mem...

Page 112: ...Total Continuous display of the totalising memory Last item Continuous display of the last item ORG data database Continuous display of the organizational data memory Fixed values checking Continuous display of the activated setpoint memory Show barcode Continuous display of the scanned barcode Code128 Weight value Display bar code128 with the current stability weight value or counter value Factor...

Page 113: ...le mechanically damaged apply preload readjust replace load cell 51 Value above zero range limit Measured value greater than stored zero point scale mechanically damaged remove preload readjust replace load cell 52 Overload Measured value across weighing range unload scale defective load cell 53 Underload Turn on scale with preload and then remove preload switch off display on the keypad restore w...

Page 114: ...114 ...

Page 115: ...115 ...