Issue 6 (February 2005)
Technical Ser vice Manual
S m i t h s M e d i c a l
6 — 4
M S 1 6 A S y r i n g e D r i v e r
6
Battery replacement
The unit uses an IEC 6LR61 battery (9 V, alkaline, PP3 size),
e.g. DURACELL MN1604.
The use of rechargeble cells are
not
recommended.
Some Duracell type batteries do not fit snugly in the battery
conmpartment. The metal can on the battery catches and deforms at
th rear of the compartment.
A new battery must be fitted when the indicator lamp fails to flash.
1.
Remove the driver from the lockbox (if fitted). See
Chapter 4,
Disassembly and Assembly.
2.
Slide off the battery compartment cover at the back of the driver.
+
_
GM0305-A
Figure 6-1 Battery replacement
3.
Turn the driver upside down and carefully remove and discard
the battery (ensure that national or local regulations concerning
battery disposal are followed.
4.
Insert a new battery into the battery compartment, ensuring
correct polarity as shown by the illustration in the battery
compartment.
Notes
:
The driver is protected against the effects of accidental
reverse polarity of the battery.
The alarm sounds for about 15 seconds after the battery
is fitted.
If fitting the battery is difficult, insert the rear of the
battery first and then push the terminal end down.
If the fit is still tight, source an alternative brand of
6LR61 alkaline battery.
6.
Fit the battery compartment cover, ensuring that it snaps into
place when closed
7.
Place the driver in the lockbox (if fitted). See
Chapter 4,
Disassembly and Assembly
.
SM0105-6-6.PMD
2/23/2005, 3:03 PM
4