Te c h n i c a l Ser vice Manual
I s s u e 6 ( February 2 0 0 5 )
4 — 9
S m i t h s M e d i c a l
4
M S 1 6 A S y r i n g e D r i v e r
Note
:
The motor case has two pegs on the front face, align the
motor so that the pegs do not foul the locating pillar in the
case assembly.
h. Assemble the actuator and back bearing and the leadscrew and
bearing.
3. Actuator and bearing
a.
Ensure that the motor clamp (item 13) is loose.
b.
Depress the plunger into the half nut and slide the actuator
assembly onto the leadscrew from the back bearing end.
Figure 4-1
shows the correct way the actuator (item 18) faces.
c.
Fit the back bearing to the end of the leadscrew.
d. Assemble the leadscrew and bearing assembly.
4. Leadscrew and bearing assembly
a.
Align the pin on the leadscrew and bearing assembly with the
slot in the cam assembly and press the leadscrew into position.
b.
Ensure that both front and back bearings fit snugly in their
recess in the case assembly.
c.
Check the motor is still correctly positioned (see
Figure 4-5)
. Use
a fresh adhesive pad on the clamp to hold the wires.
Figure 4-5 Piezo-transducer wiring
d. Tighten the clamp screw.
e.
Assemble the case.
GM1196-D
Position sounder leads
on self adhesive pad
Insulation
SM0105-4-6.PMD
2/23/2005, 2:56 PM