Te c h n i c a l S e r v i c e M a n u a l
I s s u e 6 ( February 2 0 0 5 )
2 — 3
S m i t h s M e d i c a l
M S 1 6 A S y r i n g e D r i v e r
2
Chapter 2 - Specification and Standards
Type:
Syringe driver with motor driven linear actuator, pulsed motion. Internal 9 V
alkaline battery power source. Digital electronic rate control.
Automatic switch off when the syringe is empty.
Infusion time:
< 30 min. - 60 hours
Rate range:
Variable delivery rate of 0 - 99 mm/h in steps of 1 mm/h.
Motor operating interval:
420 ÷ rate = seconds
Drive accuracy:
The accuracy of the linear drive mechanism that pushes the syringe plunger
forward: ± 5%
Drive force:
The force the linear drive mechanism that pushes the syringe plunger forward can
produce: >3 Kg
Actuator movement:
0.12 mm every time motor turns.
Actuator stroke:
60 mm available (depends on syringe).
Syringe sizes:
2 ml to 35 ml
The holster supplied will enclose most types of syringe up to 20 ml capacity.
Occlusion pressure:
3 - 5 Kg
Maximum actuator force 50 N (5 Kgf) @ 9 V
Controls:
START/TEST
and
mm PER
rate adjusters (tens and units digit controls).
Alarm:
Audible frequency:
3 kHz ± 0.5 kHz (end of travel: occlusion).
Duration:
5 - 20 secs.
Volume:
Audible at 1m with normal or corrected hearing
SM0105-2-6.PMD
2/23/2005, 2:55 PM
3