Te c h n i c a l S e r v i c e M a n u a l
I s s u e 6 ( February 2 0 0 5 )
5 — 7
S m i t h s M e d i c a l
M S 1 6 A S y r i n g e D r i v e r
5
Figure 5-4 Using the cam adjustment tool
2. 9 Volt thrust test
a.
Fit the dummy case to the driver.
b.
Ensure that the battery and battery cover are removed.
c.
Load the unit onto the Thrust rig as shown in
Figure 5-5 Thrust
rig set-up
. Set the actuator 50 mm from end of travel.
Syringe Driver
Actuator assembly
GM1276-D
Thrust test rig
Figure 5-5 Thrust rig set-up
SM0105-5-6.PMD
2/23/2005, 3:07 PM